Procedural and Hemodynamic Outcomes of Contemporary Self-Expanding Valves for Transcatheter Aortic Valve Replacement: A Multicenter Comparison From the AMTRAC Registry.
Amos Levi, Yeela Talmor-Barkan, Marco Barbanti, Sofia Sammartino, Giuliano Costa, Carmelo Grasso, Ole De Backer, Davorka Lulic,
Department of Cardiology, Rabin Medical Centre, Petach-Tikva, Israel. · The Department of Medicine and Surgery, Università degli Studi di Enna Kore, Enna, Italy.
Key points
From the paper's conclusion- Contemporary self-expanding platforms demonstrate comparable 1-year mortality and stroke rates. Periprocedural vascular and bleeding safety profiles are strongly modulated by center-specific clustering and practice variations. Conversely, the superior hemodynamic performance of the Evolut platform remains independent of institutional factors. These findings support an anatomy-driven approach to valve selection integrating patient profiles with institutional practices.
Abstract
BACKGROUND: Comparative data among contemporary self-expanding transcatheter heart valve designs remain limited. We compared clinical and hemodynamic outcomes following transcatheter aortic valve replacement using 2 leading contemporary platforms. METHODS: This multicenter cohort study utilized the AMTRAC registry to evaluate consecutive patients treated with either Evolut PRO/PRO+/FX or Navitor. The primary analysis employed a 1:1 propensity score-matched cohort, utilizing multilevel mixed-effects models to adjust for center-level clustering. The primary endpoint was 1-year all-cause mortality. RESULTS: Of 2673 patients, propensity score matching yielded a balanced cohort of 1200 patients. At 1-year, all-cause mortality was similar between groups (Navitor: 7.1% vs. Evolut: 9.2%; hazard ratio 0.75; 95% CI: 0.47-1.21; p = 0.24). In-hospital mortality (0.8%), periprocedural stroke (2.3%), permanent pacemaker implantation (Navitor: 14.4% vs. Evolut: 15.9%; p = 0.554), and paravalvular leak (p = 0.221) were comparable. Initial differences showing lower vascular and bleeding events with Navitor were completely attenuated after accounting for center-level clustering (vascular: p = 0.838; bleeding: p = 0.745). Conversely, postprocedural mean transaortic gradient was lower with Evolut (7.3 ± 3.9 mmHg vs. 8.6 ± 4.1 mmHg; p < 0.001), a hemodynamic advantage that remained highly robust after center adjustment (odds ratio 2.43; 95% CI: 1.66-3.58; p < 0.001). Sensitivity analyses yielded consistent results. CONCLUSIONS: Contemporary self-expanding platforms demonstrate comparable 1-year mortality and stroke rates. Periprocedural vascular and bleeding safety profiles are strongly modulated by center-specific clustering and practice variations. Conversely, the superior hemodynamic performance of the Evolut platform remains independent of institutional factors. These findings support an anatomy-driven approach to valve selection integrating patient profiles with institutional practices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pooled Periprocedural and Short-Term Outcomes After Transcatheter Aortic Valve Replacement With Contemporary Self-Expanding Valves: A Systematic Review and Meta-Analysis.Structural heart : the journal of the Heart Team · 2026
- Conduction Disturbances and Permanent Pacemaker Implantation After TAVR With Self-Expanding Devices: The CONNAVIPRO Study.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Early Post-transcatheter Aortic Valve Implantation (TAVI) Haemodynamics With SAPIEN Balloon-Expandable and Evolut Self-Expanding Valve Platforms: A Retrospective Single-Centre Cohort Study.Cureus · 2026
- Real-World Comparison of Balloon-Expandable Valves for Transcatheter Aortic Valve Replacement: Data From the Aortic+Mitral TRAansCatheter Valve Registry.Structural heart : the journal of the Heart Team · 2026
- Stroke After Transcatheter Aortic Valve Replacement With Contemporary Valves: Insights From the OCEAN-TAVI Registry.Structural heart : the journal of the Heart Team · 2026
- Current Generation Self-Expanding and Balloon-Expandable Transcatheter Aortic Valve Replacement: A Retrospective Multicenter Analysis.The Annals of thoracic surgery · 2026