Dual-mode colorimetric sensing of bilirubin using a PDDA@peroxomolybdate composite: A spectrophotometric platform and cost-effective cotton-swab sensor for point-of-care testing.

Asiye Aslıhan Avan, Hayati Filik, Mine Kucur, Neslihan Cabıoğlu

Istanbul University-Cerrahpaşa, Faculty of Engineering, Department of Chemistry, 34320 Avcılar, Istanbul, Turkiye. · Istanbul University-Cerrahpaşa, Faculty of Engineering, Department of Chemistry, 34320 Avcılar, Istanbul, Turkiye. Electronic address: filik@iuc.edu.tr.

Key points

From the paper's conclusion
  • This dual-mode sensing approach offers flexibility by letting users choose between the high analytical accuracy of laboratory instrumentation and the convenience of wide-range, on-site screening, depending on clinical requirements.

Abstract

In this work, the synthesis and characterization of a poly(diallyldimethylammonium chloride) (PDDA)@peroxomolybdate composite are presented for the development of a dual-mode colorimetric sensing platform for point-of-care bilirubin (BLR) detection. The platform combines a laboratory-based spectrophotometric assay with a smartphone-assisted cotton-swab sensor, enabling versatile and accessible screening. The spectrophotometric mode exhibited a wide linear detection range of 0.1-4.5 mg/dL (1.72-77.0 μM) and a limit of detection (LOD) of 0.032 mg/dL (0.54 μM). In parallel, the smartphone-based assay, utilizing normalized red-channel intensity values, demonstrated a broader linear range of 0.1-9.0 mg/dL (1.72-154 μM) with an LOD of 0.046 mg/dL (0.78 μM). This dual-mode sensing approach offers flexibility by letting users choose between the high analytical accuracy of laboratory instrumentation and the convenience of wide-range, on-site screening, depending on clinical requirements. Furthermore, the selectivity of the platform was comprehensively evaluated, and its practical applicability was successfully validated using adult and neonatal blood serum samples. The results highlight the strong potential of the proposed platform for reliable clinical monitoring of BLR levels.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers