Spectrochimica acta. Part A, Molecular and biomolecular spectroscopy

High-sensitivity label-free terahertz metasurface sensor for REGγ detection in complex biofluids.

Chuhua Huang, Hui Chen, Ming Li, Yufan Zhang, Fangrong Hu, Dongxia Li, Lei Li

Guangxi Key laboratory of Optoelectronic Information Processing, Guilin University of Electronic Technology, Guilin, 541004, China; College of Metrology Measurement and Instrument, China Jiliang University, Hangzhou, Zhejiang, 310018, China. · Department of Trauma-Emergency & Critical Care Medicine, Shanghai Fifth People's Hospital, Fudan University, Shanghai, 200240, China; Joint Center for Translational Medicine, Shanghai Fifth People's Hospital, Fudan University and School of Life Science, East China Normal University, Shanghai, 200240, China.

Key points

From the paper's conclusion
  • Here, we developed a label-free terahertz (THz) metasurface sensor with an HS-PEG-NHS-functionalized biointerface for high-sensitivity detection of REGγ.

Abstract

Accurate detection of ultralow-abundance regulatory proteins in biofluids is essential for disease diagnosis and clinical analysis. REGγ (PA28γ/PSME3), an ATP-independent proteasome activator involved in protein degradation, cell-cycle regulation, and bone metabolism, remains challenging to quantify due to its low serum concentration and interference from complex biological matrices, which limit the sensitivity and reliability of conventional immunoassays. Here, we developed a label-free terahertz (THz) metasurface sensor with an HS-PEG-NHS-functionalized biointerface for high-sensitivity detection of REGγ. The engineered interface facilitates covalent antibody immobilization and provides a PEG-based functional interface for REGγ recognition, thereby supporting sensitive detection in complex biological samples. The sensor exhibits a sensitivity of 75 GHz RIU-1 and shows a measurable response at the lowest tested REGγ concentration of 50 pg mL-1 over a broad linear range. Furthermore, the platform demonstrates enabled detection of REGγ in clinical serum samples, with the lowest REGγ concentration measured among the analyzed clinical serum samples being 90.77 pg mL-1 and showing distinguishable responses between the limited osteoporosis and healthy control samples examined in this preliminary study. This work provides a simple, reliable, and label-free THz biosensing strategy for sensitive detection of low-abundance protein biomarkers in complex biological samples.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

High-sensitivity label-free terahertz metasurface sensor for REGγ detection in complex biofluids — Science Skim