Design of the Optical System for a grating front-dispersive near-infrared spectrometer with a large-spot.
Guimin Cai, Jiachen Li, Junhui Li
Shanghai Lengguang Technology Co., Ltd., Shanghai, China. · College of Information and Electrical Engineering, China Agricultural University, Beijing, China.
Key points
From the paper's conclusion- The results demonstrated that the system maintained acceptable optical throughput while providing good spectral repeatability, with absorbance noise as low as 0.0006 Abs.
Abstract
Classical grating front-dispersive near-infrared (NIR) spectrometers, constrained by their small illumination spot size (typically around 1 cm × 2 cm), often require destructive sample pretreatment and repeated measurements at multiple positions to ensure data representativeness when analyzing nonuniform samples. These limitations significantly increase the overall measurement time. To address this issue, we designed a grating front-dispersive large-spot NIR spectrometer optical system, which enables rapid and non-destructive detection of nonuniform samples by increasing the illumination spot size. The proposed system integrated a tungsten halogen light source and employed a cross-asymmetric Czerny-Turner optical configuration. A concave mirror was used to expand the beam to form a 4 cm × 2 cm illumination spot, while an integrating sphere and an InGaAs detector were used to uniformly collect and detect the diffuse-reflection signals. Key parameters of the integrating sphere, such as height and port size, were optimized through optical simulations using the LightTools platform and further validated through experimental testing to determine the optimal sizes. The results demonstrated that the system maintained acceptable optical throughput while providing good spectral repeatability, with absorbance noise as low as 0.0006 Abs. The system was successfully applied to the non-destructive discrimination of tobacco leaves from the upper, middle, and lower stalk positions. A partial least squares discriminant analysis (PLS-DA) model exhibited outstanding classification performance, achieving high classification correct rate and rational recognition rate. This design provides a reliable technical solution for the non-destructive detection and real-time monitoring of nonuniform samples in agriculture.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative study of aperture-induced aberrations in reflective and transmissive optical spectrometers using ZEMAX.Journal of the Optical Society of America. A, Optics, image science, and vision · 2026
- Design and Demonstration of Compact and Lightweight Imaging Spectrometer Based on Schwarzschild Reflector Systems Using Commercial Off-the-Shelf OpticsSensors · 2025
- Optical Calibration Method and Calibration Test Analysis of Infrared Fourier Transform Spectrometer used in AIMSResearch in Astronomy and Astrophysics · 2026
- Dual-probe confocal reflectance microscope for nondestructive spectral reflectance measurements from both sides of coated optical components.Applied optics · 2026
- Continuously tunable attenuation stage for weak-light radiance calibration using an integrating sphere and a dual-aperture tube.Applied optics · 2026
- Sample-Specific Signal-to-Noise Ratio in Benchtop and Handheld NIR Reflectance SpectroscopyACS Measurement Science Au · 2026