Optical Calibration Method and Calibration Test Analysis of Infrared Fourier Transform Spectrometer used in AIMS
Lianwei Zhao, Zhanhu Wang, Meng Mou, Ang Jin, Jianwen Hua, Lei Ding, Yuanyong Deng, Dong-Guang Wang,
Key points
From the paper's conclusion- The study named Accurate Infrared Magnetic Field Measurements of the Sun (AIMS) is an observational instrument composed of a telescope, the Ultra-High Spectral Resolution Imaging Infrared Fourier Transform Spectrometer (FTIR), a polarization system and other scientific instrument, which is designed for measuring the solar magnetic field.
Abstract
The study named Accurate Infrared Magnetic Field Measurements of the Sun (AIMS) is an observational instrument composed of a telescope, the Ultra-High Spectral Resolution Imaging Infrared Fourier Transform Spectrometer (FTIR), a polarization system and other scientific instrument, which is designed for measuring the solar magnetic field. For FTIR system, the registration accuracy between the optical axis of the interferometer and that of the subsequent imaging optics significantly affects the overall instrument performance. When the FTIR system employs a cryogenic optical design, achieving high-precision optical axis alignment becomes considerably more difficult. Based on the FTIR developed in AIMS, this paper presents a high precision optical axis registration method through spectral calibration of the spectrometer. In this method, the optical axis is evaluated by comparing the deviation between the instrument line shape (ILS) and the theoretical curve. The proposed method is particularly applicable to spectrometers with high spectral resolution and a large number of detector pixels. Owing to the high spectral resolution and the large optical path difference of the FTIR, speckle from the light source significantly distorts the ILS. Conventional spectral calibration methods, such as using an integrating sphere to generate a uniform area light source, are incapable of realizing effective spectral calibration. A point-source spectral calibration method is proposed in this paper, and successfully applied to the spectral calibration of the FTIR, obtaining a spectrum with spectral resolution of 0.004cm−1 over the full field of view.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The Pointing and Tracking System of AIMSResearch in Astronomy and Astrophysics · 2026
- The overview of the Infrared Fourier Transform Spectrometer used in AIMSResearch in Astronomy and Astrophysics · 2026
- A Data Processing Method for AIMS Imaging FTIRResearch in Astronomy and Astrophysics · 2026
- On-site optical wavefront testing and control for the AIMS solar telescopeResearch in Astronomy and Astrophysics · 2026
- Total power calibration in FTIR emission spectroscopy for finite interferogramsAtmospheric Measurement Techniques · 2026
- Dynamic Optical Alignment Method for AIMSResearch in Astronomy and Astrophysics · 2026