Quantitative gamma-based evaluation of a novel four-dimensional computed tomography reconstruction algorithm for stack artifact reduction.
M Ahmad, J Dickmann, N A Lackner, O J Ott, R Fietkau, S Corradini, C Bert, J Szkitsak
Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg(FAU), Erlangen, Germany. · Siemens Healthineers AG, Forchheim, Germany.
Key points
From the paper's conclusion- Gamma-based evaluation provided an objective method to detect and localize reconstruction differences in 4DCT, supporting transparent clinical implementation of advanced reconstruction algorithms in radiotherapy.
Abstract
BACKGROUND AND PURPOSE: Four-dimensional computed tomography (4DCT) is essential for motion-resolved imaging in radiotherapy, but stack alignment artifacts caused by irregular breathing may compromise anatomical accuracy. This study established a quantitative framework to evaluate reconstruction differences between clinical reference reconstruction (CRR) and stack artifact-reduced reconstruction (SARR) and identify clinically relevant changes. MATERIALS AND METHODS: Retrospective 4DCT datasets from 80 lung cancer patients were analyzed, comprising a rated cohort (n = 30) and an independent control cohort (n = 50). Geometric, intensity-based, and combined metrics, including gamma analysis, were calculated across ten respiratory phases for segmented organs-at-risk. Correlations with reader ratings identified metrics that best reflected clinically relevant reconstruction differences. RESULTS: Most similarity and intensity-based measures showed limited association with image quality differences (Pearson correlation coefficient r < 0.5). Gamma-based metrics demonstrated the strongest correlation, with relative gamma area > 0.3 reaching r ≈ 0.59 for phase-based data (95% confidence interval 0.30-0.79). Reconstruction differences were predominantly observed in the liver and lower lung lobes, consistent with respiratory motion, while the spinal cord showed minimal variation. Robust z-score thresholding enabled automated categorization of moderate and substantial deviations. In the control cohort, 12% showed substantial and 14% moderate changes for phase-based reconstructions. In the patient with the largest changes, the liver appeared as a double contour in the CRR and as a single contour in the SARR. CONCLUSIONS: Gamma-based evaluation provided an objective method to detect and localize reconstruction differences in 4DCT, supporting transparent clinical implementation of advanced reconstruction algorithms in radiotherapy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- New reconstruction algorithm for stack artifact reduction in breathing-adaptive four-dimensional computed tomography.Physics and imaging in radiation oncology · 2026
- A clinical evaluation of spiral 4DCT and intelligent 4DCT sequence scanning.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- Task-based image quality and dosimetric validation of HyperSight CBCT for adaptive radiotherapy.Journal of applied clinical medical physics · 2026
- A clinical evaluation of HyperSight CBCT enhanced offline adaptive breast radiotherapy.Frontiers in oncology · 2026
- Robustness of deep learning-based denoising 4DCBCT methods.Physics in medicine and biology · 2026
- Towards adaptive radiotherapy: Comparing HyperSight and standard CBCT reconstructions on a C-Arm Linac.Journal of applied clinical medical physics · 2026