Quantitative gamma-based evaluation of a novel four-dimensional computed tomography reconstruction algorithm for stack artifact reduction.

M Ahmad, J Dickmann, N A Lackner, O J Ott, R Fietkau, S Corradini, C Bert, J Szkitsak

Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg(FAU), Erlangen, Germany. · Siemens Healthineers AG, Forchheim, Germany.

Computational

Key points

From the paper's conclusion
  • Gamma-based evaluation provided an objective method to detect and localize reconstruction differences in 4DCT, supporting transparent clinical implementation of advanced reconstruction algorithms in radiotherapy.

Abstract

BACKGROUND AND PURPOSE: Four-dimensional computed tomography (4DCT) is essential for motion-resolved imaging in radiotherapy, but stack alignment artifacts caused by irregular breathing may compromise anatomical accuracy. This study established a quantitative framework to evaluate reconstruction differences between clinical reference reconstruction (CRR) and stack artifact-reduced reconstruction (SARR) and identify clinically relevant changes. MATERIALS AND METHODS: Retrospective 4DCT datasets from 80 lung cancer patients were analyzed, comprising a rated cohort (n = 30) and an independent control cohort (n = 50). Geometric, intensity-based, and combined metrics, including gamma analysis, were calculated across ten respiratory phases for segmented organs-at-risk. Correlations with reader ratings identified metrics that best reflected clinically relevant reconstruction differences. RESULTS: Most similarity and intensity-based measures showed limited association with image quality differences (Pearson correlation coefficient r < 0.5). Gamma-based metrics demonstrated the strongest correlation, with relative gamma area > 0.3 reaching r ≈ 0.59 for phase-based data (95% confidence interval 0.30-0.79). Reconstruction differences were predominantly observed in the liver and lower lung lobes, consistent with respiratory motion, while the spinal cord showed minimal variation. Robust z-score thresholding enabled automated categorization of moderate and substantial deviations. In the control cohort, 12% showed substantial and 14% moderate changes for phase-based reconstructions. In the patient with the largest changes, the liver appeared as a double contour in the CRR and as a single contour in the SARR. CONCLUSIONS: Gamma-based evaluation provided an objective method to detect and localize reconstruction differences in 4DCT, supporting transparent clinical implementation of advanced reconstruction algorithms in radiotherapy.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Quantitative gamma-based evaluation of a novel four-dimensional computed tomography reconstruction algorithm for stack artifact reduction — Science Skim