Physics and imaging in radiation oncology

Functional lung avoidance planning with combined ventilation-perfusion imaging versus individual functional imaging in lung cancer radiotherapy.

Genta Michimata, Yujiro Nakajima, Tomoki Kimura, Nobuki Imano, Shinya Ito, Yukio Fujita, Naoki Tohyama

Department of Radiological Sciences, Komazawa University, 1-23-1 Komazawa, Setagaya-ku, Tokyo, 154-8525, Japan. · Department of Radiation Oncology, Kochi Medical School, Kochi University, Kohasu, Oko-cho, Nankoku-shi, Kochi 783-8505, Japan.

Key points

From the paper's conclusion
  • Integrating 4DCT ventilation and SPECT perfusion into radiotherapy planning improved functional lung sparing compared with anatomical or single-modality planning. Tumor proximity to functional lung influenced the magnitude of dose-volume benefits.

Abstract

BACKGROUND AND PURPOSE: Functional lung avoidance radiotherapy using four-dimensional computed tomography (4DCT) ventilation or single-photon emission computed tomography (SPECT) perfusion has been studied independently, however few studies have evaluated their combined use. The aim of this study was to examine whether integrating both modalities improved sparing of high functional lung (HFL) in non-small cell lung cancer (NSCLC). MATERIALS AND METHODS: Forty-seven patients with NSCLC underwent 4DCT ventilation and SPECT perfusion imaging. One patient was excluded, leaving 46 for analysis. Functional lung maps were segmented into high, middle, and low functional lung. Four volumetric-modulated arc therapy (VMAT) plans using 6 MV photons per patient were generated: anatomical (A), ventilation-guided (V), perfusion-guided (Q), and combined ventilation-perfusion-guided (VQ). Dose-volume and dose-function metrics were compared among plans. Correlations between tumor proximity to HFL and dose reduction were assessed. RESULTS: All function-guided plans significantly reduced HFL dose compared with A plans (p < 0.001). Among the function-guided plans, VQ plans yielded the lowest mean lung doses (MLDs) in high ventilated lung (VQ plans, 11.5 Gy; A plans, 12.4 Gy) and high perfused lung (VQ plans, 10.0 Gy; A plans, 11.1 Gy). Function-weighted metrics, including functional MLD, were also lowest with VQ plans, indicating improved functional lung sparing. Dose reduction correlated with tumor proximity to HFL, with the strongest associations observed at planning target volume expansions between 25 and 45 mm. CONCLUSIONS: Integrating 4DCT ventilation and SPECT perfusion into radiotherapy planning improved functional lung sparing compared with anatomical or single-modality planning. Tumor proximity to functional lung influenced the magnitude of dose-volume benefits.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers