Interplay effect evaluation on lung and esophageal cancer patients treated with compact proton system in free breathing.
Giorgio Cartechini, Esther Kneepkens, Gloria Vilches-Freixas, Richard Canters, Mirko Unipan, Marije Velders, Indra Lubken, Maud Cobben,
Maastro, Maastricht Radiation Oncology Institute, Maastricht, the Netherlands.
Key points
From the paper's conclusion- Internal target volume-based robust optimization combined with machine-specific beam delivery characteristics provides motion robustness for most lung and esophageal patients treated in free-breathing. Additional motion mitigation was not required for most patients treated under conventional fractionation. However, small targets with large motion remain at risk of interplay-induced underdosage, warranting caution, as this combination may compromise plan robustness.
Abstract
BACKGROUND AND PURPOSE: Pencil beam scanning proton therapy is sensitive to respiratory motion, which can cause dose discrepancies due to the interplay between tumor and beam delivery motion. This study evaluates the clinical impact of interplay for lung and esophageal cancer patients treated in free-breathing, without rescanning. MATERIALS AND METHODS: Interplay was evaluated in 36 patients (18 lung, 18 esophageal) using both predictive and a posteriori approaches. The predictive method, for pre-treatment assessment, employed 24 synthetic sinusoidal breathing patterns to model respiratory motion. The a posteriori method incorporated patient-specific breathing traces and machine log files. Fraction-accumulated dynamic 4D dose distributions were generated and compared with the institutional 4D robustness evaluation. RESULTS: The predictive method agreed closely with a posteriori evaluation, with a median Relative Distribution Error < 1% across all dose-volume metrics. Interplay-induced underdosage in the first fraction (-1% to -2%) decreased after ∼5 fractions, with cumulative D98%, D95% and V95% values converging to within 2% of planned doses. Two plans, one lung and one esophageal, showed persistent underdosage exceeding 5%, consistent with 4D robustness evaluation failures. For the esophageal case, a revised beam arrangement restored robustness. CONCLUSIONS: Internal target volume-based robust optimization combined with machine-specific beam delivery characteristics provides motion robustness for most lung and esophageal patients treated in free-breathing. Additional motion mitigation was not required for most patients treated under conventional fractionation. However, small targets with large motion remain at risk of interplay-induced underdosage, warranting caution, as this combination may compromise plan robustness.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Impact of accumulated target dose on tumor response in distal esophageal cancer treated with intensity-modulated proton therapy.Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2026
- Impact of anatomical changes on simulated proton arc therapy and intensity‐modulated proton therapy for lung cancerJournal of Applied Clinical Medical Physics · 2026
- Trigger-based adaptive proton therapy and changes in target coverage and cardiac dose in esophageal cancer.Physics and imaging in radiation oncology · 2026
- Breathing-phase selection for gated proton therapy in lung cancer towards temporally targeted ultra-high dose rate delivery.Physics and imaging in radiation oncology · 2026
- Phase 1 Study of Online Proton Adaptive RadioTherapY (PARTY).International journal of radiation oncology, biology, physics · 2026
- Machine Learning-Based Beam Delivery Time Model for Mevion S250i With Hyperscan TechnologyInternational Journal of Particle Therapy · 2026