Manual Instrumentation Is Associated With Greater Precision Than Patient-Specific Instrumentation For Initial Tibial Resections In Unrestricted Kinematic Alignment Total Knee Arthroplasty: A Comparative Cohort Study.
Charles Rivière, Giuseppe Aloisi, Loic Villet
Bordeaux Arthroplasty Research Institute (BARI), 6 rue Georges Negrevergne, 33700 Mérignac, France; Clinique du Sport Bordeaux Mérignac, 4 rue Georges Negrevergne, 33700 Mérignac, France. Electronic address: charles.riviere@bari-arthroplasty.com. · Bordeaux Arthroplasty Research Institute (BARI), 6 rue Georges Negrevergne, 33700 Mérignac, France; Università degli studi dell'Aquila, Department of Life, Health and Environmental Sciences, 67100 L'Aquila, Italy.
Key points
From the paper's conclusion- Manual instrumentation and PSI achieved similar mean accuracy for initial tibial resections in uKA-TKA. However, manual instrumentation was associated, in this two-surgeon series, with greater medial reproducibility and fewer large resection deviations, with a numerically lower, non-significant recut rate. Within the sequential uKA workflow - including caliper verification and corrective recut when required - both techniques allow reliable convergence toward the predefined resurfacing target.
Abstract
PURPOSE: Accurate tibial resection is critical in unrestricted kinematic alignment total knee arthroplasty (uKA-TKA), as component positioning aims to restore the native joint line through anatomical resurfacing. This study compared the reproducibility of initial tibial resections performed with manual instrumentation versus patient-specific instrumentation (PSI) in uKA-TKA. The secondary objective was to compare tibial recut rates between techniques. METHODS: This retrospective comparative cohort study analyzed prospectively collected intraoperative measurements from 124 primary uKA-TKAs performed between June 2024 and January 2026. Tibial resections were performed using manual instrumentation in 83 cases and PSI in 41 cases. Target and measured tibial resection thicknesses were recorded intraoperatively at the medial (MTP) and lateral tibial plateaus (LTP) using an analog sliding caliper (0.5-mm resolution). Resection error (Δ) was defined as measured minus target thickness. Accuracy was expressed as mean signed error and precision as the standard deviation of signed error. Absolute error thresholds and tibial recut rates were also analyzed. RESULTS: Mean signed error did not differ significantly between techniques for either plateau (medial: 0.15 mm PSI vs -0.10 mm manual, p = 0.158; lateral: 0.10 mm PSI vs -0.05 mm manual, p = 0.378), indicating comparable average accuracy. However, medial precision was significantly greater with manual instrumentation (SD 0.64 mm vs 1.02 mm with PSI, p = 0.002), with lower mean absolute error (0.38 mm vs 0.73 mm; p = 0.007). Fewer medial resections exceeded the ±1.5 mm threshold in the manual group (2.4% vs 9.8%, p = 0.024). No significant differences were observed laterally. Angular tibial recut was performed in 12.0% of manual cases and 22.0% of PSI cases (p = 0.187). CONCLUSION: Manual instrumentation and PSI achieved similar mean accuracy for initial tibial resections in uKA-TKA. However, manual instrumentation was associated, in this two-surgeon series, with greater medial reproducibility and fewer large resection deviations, with a numerically lower, non-significant recut rate. Within the sequential uKA workflow - including caliper verification and corrective recut when required - both techniques allow reliable convergence toward the predefined resurfacing target. LEVEL OF EVIDENCE: III; retrospective comparative study.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Kinematic Alignment Using Conventional Mechanical-Alignment Instruments with 3D-Printed Tibial Recut Guide Versus Manual Mechanical Alignment in Total Knee Arthroplasty: An IPTW-Adjusted Comparative Cohort Study.Journal of clinical medicine · 2026
- 3D-printed patient-specific instrumentation significantly improves the accuracy of prosthesis alignment in unicompartmental knee arthroplasty: a retrospective study.3D printing in medicine · 2026
- Restricted Kinematic Alignment Total Knee Arthroplasty Using CT-Based Patient-Specific Jigs: An Observational Feasibility Study from an Indian Cohort.Indian journal of orthopaedics · 2026
- Accuracy and Precision of Computer-assisted Navigation Compared With Patient-Specific Instrumentation and Conventional Methods in Total Ankle Arthroplasty.The Journal of the American Academy of Orthopaedic Surgeons · 2026
- Accuracy of implant positioning in total ankle arthroplasty using patient-specific instrumentation: A weightbearing CT analysis of landmark-based parameters.Foot and ankle surgery : official journal of the European Society of Foot and Ankle Surgeons · 2026
- Improved precision and Reduced Technical Errors in Total Knee Arthroplasty Using the Velys Robotic-Assisted Solution: Single-Center Comparative Study.Acta orthopaedica Belgica · 2026