Dimensional and angular agreement of LCD printed zirconia nanohybrid orthodontic brackets with CAD targets in a single build in vitro study.

Anthony Allison, Rafael Delgado-Ruiz, Duane Shank, Richard Faber

Department of Orthodontics and Pediatric Dentistry, Stony Brook University School of Dental Medicine, 101 South Drive, Stony Brook, NY 11794, United States. Electronic address: anthony.allison@stonybrookmedicine.edu. · Department of Prosthodontics and Digital Technology, Stony Brook University School of Dental Medicine, 101 South Drive, Stony Brook, NY 11794, United States. Electronic address: rafael.delgado-ruiz@stonybrookmedicine.edu.

Preclinical

Key points

From the paper's conclusion
  • In this single build completed by one operator, mean linear dimensions and slot-wall angulation were close to the CAD targets. The results describe geometric agreement of averaged measurements within that build; they do not establish measurement reliability, workflow reproducibility, or clinical material suitability.

Abstract

INTRODUCTION: To quantify the within-build agreement of linear dimensions and slot-wall angulation of custom orthodontic brackets fabricated by LCD-masked vat photopolymerization with their nominal computer-aided design (CAD) targets. METHODS: Fifty brackets were fabricated in one build by one operator using a zirconia nanohybrid photopolymer resin (Rodin Sculpture 2.0). Two examiners measured four outcomes using ImageJ and Geomagic Control X, with three repetitions per examiner and program. Only the 12-reading mean for each specimen and outcome was retained. Means, standard deviations (SDs), confidence intervals (CIs), and conformity with investigator-defined margins (±0.05mm and ±0.35°) were summarized. These margins were not validated clinical thresholds. The brackets were technical replicates of one build; measurement reliability and between-build reproducibility could not be estimated. RESULTS: The 50 consolidated specimen-level values yielded means (SDs) of 0.557 (0.015) mm for slot height, 0.711 (0.017) mm for slot depth, 0.999 (0.006) mm for base height, and 7.032° (0.174°) for slot-wall angulation. The corresponding 95% CIs were 0.552-0.561mm, 0.706-0.716mm, 0.997-1.000mm, and 6.982°-7.081°. Specimen-level conformity with the investigator-defined limits was 94%, 94%, 100%, and 96%, respectively. CONCLUSIONS: In this single build completed by one operator, mean linear dimensions and slot-wall angulation were close to the CAD targets. The results describe geometric agreement of averaged measurements within that build; they do not establish measurement reliability, workflow reproducibility, or clinical material suitability.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers