Quantitative characterization of spatial frequency-dependent effects of conjugate gradient reconstruction in relation to k-space fill ratio in non-contrast-enhanced ultra-short echo time brain magnetic resonance angiography.
Shohei Harada, Yasuo Takatsu, Kazuhiro Murayama, Yunosuke Kumazawa, Yuichiro Sano, Masato Ikedo
Department of Radiology, Fujita Health University Hospital, 1-98, Dengakugakubo, Kutsukake-cho, Toyoake, Aichi 470-1192, Japan; Graduate School of Medical Sciences, Fujita Health University, 1-98, Dengakugakubo, Kutsukake-cho, Toyoake, Aichi 470-1192, Japan. Electronic address: shohei.harada@fujita-hu.ac.jp. · Graduate School of Medical Sciences, Fujita Health University, 1-98, Dengakugakubo, Kutsukake-cho, Toyoake, Aichi 470-1192, Japan; Department of Molecular Imaging, Clinical Collaboration Unit, School of Medical Sciences, Fujita Health University, 1-98, Dengakugakubo, Kutsukake-cho, Toyoake, Aichi 470-1192, Japan.
Key points
From the paper's conclusion- SNR increased with the k-space fill ratio and was not significantly affected by CG reconstruction.
Abstract
Ultrashort echo time magnetic resonance angiography (UTE-MRA) with radial sampling is increasingly used for non-contrast cerebrovascular imaging due to its low flow dependence and robustness to susceptibility effects. Reducing the k-space fill ratio to shorten acquisition time may alter the spatial-frequency characteristics of reconstructed images. Conjugate gradient (CG)-based iterative reconstruction compensates for incomplete k-space sampling; however, its frequency-dependent effects remain insufficiently characterized. This study evaluated the spatial-frequency characteristics of CG reconstruction in radial UTE-MRA using the normalized integrated power spectrum (NIPS). Phantom and clinical head UTE-MRA were performed at k-space fill ratios of 2.2%-35.2%, with and without CG reconstruction. NIPS findings were evaluated in relation to signal-to-noise ratio (SNR) and acquisition time. The effect of CG reconstruction was more pronounced in the low-frequency range at lower fill ratios, with significant higher NIPS values at 2.2% and 4.4% without CG reconstruction. Furthermore, no significant differences were observed among any combination within the low-frequency band when CG reconstruction was applied at fill ratios of 4.4% or higher (0.051 < P < 0.998). High-frequency NIPS values increased with the k-space fill ratio, indicating greater dependence on physical sampling density. SNR increased with the k-space fill ratio and was not significantly affected by CG reconstruction. In clinical analysis, no statistically significant differences were observed between CG reconstruction at 4.4% and conventional gridding at 8.8% in low- or high-frequency NIPS values within the M1 or M3 segments. Among evaluated conditions, 4.4% with CG reconstruction was selected as a practical acquisition condition.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantitative $T_2^*$ Fidelity in Accelerated Radial Ultrashort-Echo-Time MRI of the Lumbar Cartilage Endplate: A Synthetic Computational Validation StudyarXiv · 2026
- The Paradox of Radial MRI: When Repeated Sampling of K-Space Center Becomes a Double-Edged Sword.Magnetic resonance in medicine · 2026
- Radial Tiny Golden Angle UTE (tyGA‐UTE) for Cardiac MRI in the Small Animal ModelNMR in Biomedicine · 2026
- Quantitative Chemical Exchange Saturation Transfer Imaging With Golden‐Angle Radial k‐Space Sampling and Locally Low‐Rank ReconstructionNMR in Biomedicine · 2026
- Scan-specific k-space interpolation using deep Koopman autoencodersarXiv · 2026
- Feasibility of a 2-Minute Multi-Echo UTE Acquisition for Simultaneous CT-Like Bone-Weighted Imaging and Quantitative T2* Mapping of Short-T2 TissuemedRxiv · 2026