Reliability and validity of the My Jump Lab app for assessing vertical jump performance after anterior cruciate ligament reconstruction.

David Bruno Braga de Castro, Guilherme Antony Martins Santana, João Gabriel Barroso Carvalho, Márcio Almeida Bezerra, Gabriel Peixoto Leão Almeida

Knee and Sports Research Group, Program in Physiotherapy and Functioning, Federal University of Ceará, Fortaleza, Ceará, Brazil; Program in Physiotherapy and Functioning, Federal University of Ceará, Fortaleza, Ceará, Brazil. · Knee and Sports Research Group, Program in Physiotherapy and Functioning, Federal University of Ceará, Fortaleza, Ceará, Brazil.

Key points

From the paper's conclusion
  • My Jump Lab provides reliable and valid measures of vertical jump performance post-ACLR and may complement RTS assessment when force platforms are unavailable.

Abstract

BACKGROUND: To evaluate the test-retest reliability and validity of My Jump Lab compared with a force platform for assessing vertical jump performance after anterior cruciate ligament reconstruction (ACLR). METHODS: A total of 102 individuals, 6-24 months post-ACLR and cleared for return-to-sport (RTS) testing, performed the single-leg vertical jump (SLVJ) and single-leg drop jump (SLDJ), assessed using My Jump Lab and a force platform. Jump height derived from flight time (FT) and take-off velocity (ToV), reactive strength index (RSI) and limb symmetry index (LSI) were analyzed. Test-retest reliability, inter-instrument reliability, agreement and association, diagnostic accuracy, and association with force-platform-derived kinetic (force at zero velocity, peak landing force, and concentric net impulse) variables were assessed. RESULTS: My Jump Lab demonstrated excellent test-retest reliability (intraclass correlation coefficient (ICC) = 0.992-0.997; standard error of measurement = 0.24-0.41 cm; minimal detectable change = 0.65-1.14 cm). Associations with the force platform were strong to very strong for jump height (r = 0.786-0.987), RSI (r = 0.975), and LSI (r = 0.760-0.983). Inter-instrument reliability was excellent for FT and RSI (ICC = 0.974-0.983) and fair for ToV (ICC = 0.750). Bland-Altman analyses showed mean biases ranging from -1.21 to -0.08 cm for jump height and -0.29 cm/s for RSI. Accuracy for LSI ≥ 90% ranged from 87.3% to 94.1%. SLVJ jump height showed moderate-to-good associations with force-platform-derived kinetic (r = 0.534-0.678). CONCLUSIONS: My Jump Lab provides reliable and valid measures of vertical jump performance post-ACLR and may complement RTS assessment when force platforms are unavailable.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers