Three-dimensional surface morphology of retrieved polyethylene inserts and component alignment in failed total ankle arthroplasty: A Basic retrieval study.
Hiroaki Kurokawa, Kenya Iijima, Shinichi Kosugi, Masataka Yamamoto, Sayaka Wakiyama, Norihiro Tsujimoto, Yoshiyuki Kamatani, Yinghao Li,
Department of Orthopaedic Surgery, Nara Medical University, Nara, Japan. Electronic address: blackandriver@hotmail.co.jp. · Department of Mechanical and Aerospace Engineering, Graduate School of Science, and Technology, Tokyo University of Science, Noda, Chiba, Japan.
Key points
From the paper's conclusion- Surface deviation was greater in peripheral than central regions and, on relative assessment, in the posteromedial than anterolateral quadrant. An exploratory association between varus alignment change and posteromedial relative mean surface deviation warrants further investigation.
Abstract
BACKGROUND: Polyethylene (PE) damage is a recognized mode of failure after total ankle arthroplasty (TAA); however, little is known about its three-dimensional (3D) spatial distribution or its relationship with component alignment. METHODS: Nine PE inserts retrieved during revision surgery after fixed-bearing TAA were evaluated using a non-destructive 3D surface scanning technique. Normalized surface deviation maps quantified the magnitude and spatial distribution of surface morphological changes. Coronal and sagittal alignment of the tibial component and sagittal alignment of the talar component were assessed immediately after the primary surgery and immediately before revision. RESULTS: Surface morphological changes were greater in peripheral than central regions (P = 0.039). Quadrant analysis demonstrated regional differences (P = 0.016), with post hoc testing showing greater posteromedial than anterolateral change (adjusted P = 0.047). Radiographically, the tibial component demonstrated increased varus inclination at the prerevision assessment (P = 0.031), whereas sagittal tibial and talar component alignment remained unchanged. In an exploratory analysis of eight ankles, greater coronal tibial component alignment change toward varus was associated with greater posteromedial relative mean surface deviation (Spearman's ρ = -0.766, P = 0.027). CONCLUSION: Surface deviation was greater in peripheral than central regions and, on relative assessment, in the posteromedial than anterolateral quadrant. An exploratory association between varus alignment change and posteromedial relative mean surface deviation warrants further investigation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Longitudinal 3D Shape and Alignment Analysis of Post-traumatic Osteoarthritis After Tibial Plafond Fractures.Foot & ankle orthopaedics · 2026
- Association of Component Sizing on Total Ankle Arthroplasty Survivorship.The Journal of foot and ankle surgery : official publication of the American College of Foot and Ankle Surgeons · 2026
- Tibial Component Malalignment in Total Ankle Arthroplasty.Medicina (Kaunas, Lithuania) · 2026
- In vivo quantification of polyethylene wear in total ankle arthroplasty : a longitudinal study using weightbearing CT.Bone & joint research · 2026
- The effect of restoring the native ankle on clinical outcome and range of motion in total ankle arthroplasty.Foot and ankle surgery : official journal of the European Society of Foot and Ankle Surgeons · 2026
- Reaming-induced joint surface destruction in tibiotalocalcaneal nailing: a cadaveric comparison of straight and valgus-curved nails.Journal of orthopaedic surgery and research · 2026