Knot Security Under Simulated Oral Conditions: The Effects of Suture Material, Knot Configuration, and Number of Throws.
Zehra Gülerol Aydin, Batuhan Aydin, Serkan Polat
Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Bolu Abant İzzet Baysal University, Bolu, Turkey. Electronic address: zehra.gulerol@ibu.edu.tr. · Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Bolu Abant İzzet Baysal University, Bolu, Turkey. Electronic address: batuhan.aydin@ibu.edu.tr.
Key points
From the paper's conclusion- Number of throws was more strongly associated with maximum force and knot security than knot configuration. 2-throw knots were more susceptible to slippage, whereas 3- and 4-throw knots did not differ significantly. Polat's knot showed lower slippage odds in the 3-throw model but no consistent overall biomechanical superiority.
Abstract
OBJECTIVE: To compare Polat's knot, a modified surgeon's knot, with the conventional surgeon's knot and assess factors affecting knot security and mechanical performance. METHODS: This in vitro study included 324 specimens assigned according to three 4-0 suture materials, two knot configurations, three numbers of throws, and two time points. Sutures were immersed in saline at 37°C for 15 minutes before knot tying, and each throw was tightened to 5 N. Day 0 specimens were tested immediately; Day 7 specimens were stored in artificial saliva at 37°C and thermocycled between 5°C and 55°C. Maximum force was analyzed using aligned rank transform ANOVA, and knot slippage failure using Firth logistic regression. RESULTS: Maximum force was affected by knot configuration (p=0.025), time point, number of throws, and suture material (all p<0.001). Knot configuration had a very small effect (η²=0.017), whereas number of throws had a larger effect (η²=0.256). 2-throw knots had higher odds of slippage failure than 4-throw knots for silk (adjusted OR=8.858), polypropylene (adjusted OR=22.900), and polyglactin 910 (adjusted OR=14.578). In the 3-throw model, the surgeon's knot had higher odds of slippage failure than Polat's knot (adjusted OR=3.560; 95% CI, 1.076-13.838; p=0.037). CONCLUSION: Number of throws was more strongly associated with maximum force and knot security than knot configuration. 2-throw knots were more susceptible to slippage, whereas 3- and 4-throw knots did not differ significantly. Polat's knot showed lower slippage odds in the 3-throw model but no consistent overall biomechanical superiority.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Poor knot security when closing running sutures with three threads.Facts, views & vision in ObGyn · 2026
- Suture Slippage Is Influenced by Suture Type and Bone Density in a Knotless Lateral Row Rotator Cuff Repair: A Biomechanical Analysis.Arthroscopy, sports medicine, and rehabilitation · 2026
- Biomechanical comparison of suture materials across a bone tunnel edge.The Journal of hand surgery, European volume · 2026
- A Simple Knot Technique to Simultaneously Prevent Medial-Row Suture Loosening and Lateral-Row Anchor Failure in Double-Row Rotator Cuff Repair.Arthroscopy techniques · 2026
- Knot-Dependent Constructs Show Narrowly Better Ultimate Failure Loads Than Knotless Constructs in Proximal Biceps Tenodesis: A Systematic Review and Meta-analysis.Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association · 2026
- Biomechanical Analysis of Rotator Cuff Repair: Comparison of Sutures and Fixation Techniques in a Synthetic Model.Cureus · 2026