Journal of the mechanical behavior of biomedical materials

A fractographic and numerical assessment of dental zirconia implant fractures.

U Lohbauer, R Belli, U Schepke, J H W de Beus, M S Cune, S S Scherrer, C Schulbert, N Pechinger,

Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Research Laboratory for Dental Biomaterials, Department of Operative Dentistry and Periodontology, Erlangen, Germany. Electronic address: ulrich.lohbauer@fau.de. · Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Research Laboratory for Dental Biomaterials, Department of Operative Dentistry and Periodontology, Erlangen, Germany. Electronic address: renan.belli@fau.de.

Key points

From the paper's conclusion
  • Implant fractures resulted from manufacturing defects interacting with a stress-concentrating thread geometry, exacerbated by marginal bone loss and off-axis loading. These findings highlight the importance of manufacturing quality, fracture-resistant implant design, and appropriate surgical and prosthetic planning.

Abstract

INTRODUCTION: This study investigated fracture mechanisms, complementing a prospective clinical trial reporting a 16.7% five-year fracture-related failure rate. MATERIALS AND METHODS: Four retrieved, horizontally fractured ZV3 zirconia implants were examined using scanning electron microscopy (SEM), surface profilometry, and micro-computed tomography (micro-CT). Finite element analysis (FEA) assessed stress distribution under horizontal loading, considering parametric variations of different bone level support as well as variations of occlusal contact loading angles, relative to the implant axis. RESULTS: All implants fractured palatal-to-buccal, originating at the root of the first endosseous thread. SEM revealed pressing-related volumetric defects and pronounced surface damage from milling and grinding. Mean surface roughness (Rz) was 10-20 μm, while micro-CT showed volume defect sizes peaking at approximately 30 μm and reaching 50-70 μm. FEA predicted peak stress of 750.85 MPa under a 600 N horizontal load, increasing to 3963.30 MPa with 3-mm simulated marginal bone loss under worst-case conditions. A 1-mm buccolingual crown shift towards the implant axis reduced the maximum principal stress in the root of the first endosseous thread by 123.03 MPa. DISCUSSION: Fracture mechanics calculations indicated that critical defects could substantially reduce the fracture strength, potentially allowing clinically relevant masticatory forces to initiate fracture. Thread-root stress concentration, bone loss, and off-axis loading conditions further contribute increasing the fracture susceptibility. CONCLUSIONS: Implant fractures resulted from manufacturing defects interacting with a stress-concentrating thread geometry, exacerbated by marginal bone loss and off-axis loading. These findings highlight the importance of manufacturing quality, fracture-resistant implant design, and appropriate surgical and prosthetic planning.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers