A fractographic and numerical assessment of dental zirconia implant fractures.
U Lohbauer, R Belli, U Schepke, J H W de Beus, M S Cune, S S Scherrer, C Schulbert, N Pechinger,
Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Research Laboratory for Dental Biomaterials, Department of Operative Dentistry and Periodontology, Erlangen, Germany. Electronic address: ulrich.lohbauer@fau.de. · Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Research Laboratory for Dental Biomaterials, Department of Operative Dentistry and Periodontology, Erlangen, Germany. Electronic address: renan.belli@fau.de.
Key points
From the paper's conclusion- Implant fractures resulted from manufacturing defects interacting with a stress-concentrating thread geometry, exacerbated by marginal bone loss and off-axis loading. These findings highlight the importance of manufacturing quality, fracture-resistant implant design, and appropriate surgical and prosthetic planning.
Abstract
INTRODUCTION: This study investigated fracture mechanisms, complementing a prospective clinical trial reporting a 16.7% five-year fracture-related failure rate. MATERIALS AND METHODS: Four retrieved, horizontally fractured ZV3 zirconia implants were examined using scanning electron microscopy (SEM), surface profilometry, and micro-computed tomography (micro-CT). Finite element analysis (FEA) assessed stress distribution under horizontal loading, considering parametric variations of different bone level support as well as variations of occlusal contact loading angles, relative to the implant axis. RESULTS: All implants fractured palatal-to-buccal, originating at the root of the first endosseous thread. SEM revealed pressing-related volumetric defects and pronounced surface damage from milling and grinding. Mean surface roughness (Rz) was 10-20 μm, while micro-CT showed volume defect sizes peaking at approximately 30 μm and reaching 50-70 μm. FEA predicted peak stress of 750.85 MPa under a 600 N horizontal load, increasing to 3963.30 MPa with 3-mm simulated marginal bone loss under worst-case conditions. A 1-mm buccolingual crown shift towards the implant axis reduced the maximum principal stress in the root of the first endosseous thread by 123.03 MPa. DISCUSSION: Fracture mechanics calculations indicated that critical defects could substantially reduce the fracture strength, potentially allowing clinically relevant masticatory forces to initiate fracture. Thread-root stress concentration, bone loss, and off-axis loading conditions further contribute increasing the fracture susceptibility. CONCLUSIONS: Implant fractures resulted from manufacturing defects interacting with a stress-concentrating thread geometry, exacerbated by marginal bone loss and off-axis loading. These findings highlight the importance of manufacturing quality, fracture-resistant implant design, and appropriate surgical and prosthetic planning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Fracture modelling of zirconia crowns in dental implant systems using XFEM, phase-field methods, and Weibull statistics.Dental materials : official publication of the Academy of Dental Materials · 2026
- A Three-Dimensional Finite Element Analysis to Evaluate and Compare Stress Levels and Distribution Patterns in Lithium Disilicate and Zirconia Crowns.European journal of dentistry · 2026
- Effect of Connector Cross-Sectional Geometry on Fracture Resistance of Three-Unit Monolithic Zirconia Fixed Dental Prostheses Under Vertical and Oblique Loading: An In Vitro Study.Cureus · 2026
- Comparison of Fracture Resistance of 3D Printed and Milled Zirconia Veneers and Anterior Crowns.Clinical and experimental dental research · 2026
- Biaxial Flexural Strength and Fracture Behavior of Five-Unit Monolithic Zirconia Fixed Dental Prostheses: An Experimental and Finite Element Study.Journal of functional biomaterials · 2026
- Biomechanical Behavior of Monolithic Zirconia Prosthodontic Restorations in Maxillary Posterior Teeth with Reduced Periodontium: A Finite Element Analysis.Materials (Basel, Switzerland) · 2026