Accuracy of full-arch implant scans using different horizontal scan bodies in an angulated model: An in vitro study.
Tsung-Yu Lin, Tong-Mei Wang, Li-Deh Lin
Charming Dental Clinic, Private Practice, Taipei, Taiwan. · Department of Dentistry, National Taiwan University Hospital, Taipei, Taiwan.
Key points
From the paper's conclusion- Scan body design significantly influences full-arch implant scan trueness. Although all systems performed within clinically acceptable thresholds (<31 μm), comprehensive accuracy evaluations should incorporate both RMS values and the distribution of extreme deviations to ensure predictable clinical passive fit.
Abstract
BACKGROUND/PURPOSE: Various newly developed horizontal implant scan bodies have been introduced to enhance full-arch scan accuracy. However, their performance in angulated implant scenarios remains unclear. The purpose of this study was to evaluate the accuracy of four specific horizontal scan body systems. MATERIALS AND METHODS: A stone cast containing six implant abutment analogs served as the master model. Four horizontal scan body systems (ArcB GeoMEDI, Apollo Smartflag, Scan Ladder, and ioConnect) were evaluated. Ten scans were performed per group using an intraoral scanner (Medit i900c). The mean surface deviation, root mean square (RMS) values against the reference files, and the percentage of data points within specific tolerance thresholds were calculated using Geomagic Control X. One-way ANOVA was performed to test the effect of the scan body system on accuracy outcomes (α = 0.05). RESULTS: Scan Ladder (18.42 ± 2.33 μm) and ioConnect (20.54 ± 3.32 μm) exhibited the highest trueness, significantly outperforming Apollo and ArcB (P < 0.001). Scan Ladder demonstrated the lowest RMS (22.64 ± 2.8 μm), significantly superior to all other groups (P ± 0.011), and achieved the highest concentration of highly accurate data points (<50 μm; 97.47%). While all systems showed good precision, Apollo exhibited the highest rate of extreme deviations (>100 μm; 5.05%). CONCLUSION: Scan body design significantly influences full-arch implant scan trueness. Although all systems performed within clinically acceptable thresholds (<31 μm), comprehensive accuracy evaluations should incorporate both RMS values and the distribution of extreme deviations to ensure predictable clinical passive fit.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Trueness and Precision of Three Novel Scan Body Designs with Horizontal Extensions for Complete-Arch Implant Impressions: A Comparative In Vitro Study with Intraoral Scanning and Photogrammetry Systems.Journal of functional biomaterials · 2026
- Accuracy of full-arch implant scan with Nexus IOS® scan gauges versus different conventional scan bodies: An in vitro comparative studyJournal of Dentistry · 2025
- Influence of an indexing component and intraoral scanner type on the accuracy of reverse scan body digital implant scans.The Journal of prosthetic dentistry · 2026
- Effect of scan calibrator morphologies on the accuracy of calibrated complete-arch implant intraoral scans: an in vitro study.Journal of dentistry · 2026
- Digitization of a Single Tissue-Level Implant Without Scan Body - A Feasibility Study.Journal of dentistry · 2026
- In vivo evaluation comparison of noncalibrated implant scan body systems positioned following the arch shape or toward the center of the arch.The Journal of prosthetic dentistry · 2026