Accuracy of autonomous robotic buccal alveolar bone reduction for anterior crown lengthening: An in vitro study.
Yalin Zhan, Huidan Shen, Ye Han, Yongtao Yang, Xiaorui Shi, Feng Liu
First Clinical Division, Peking University School and Hospital of Stomatology & National Center of Stomatology & National Clinical Research Center for Oral Diseases & National Engineering Research Center of Oral Biomaterials and Digital Medical Devices, Beijing, PR China; Peking University Hospital of Stomatology Sanya Division (Sanya Stomatology Center). · Peking University Hospital of Stomatology Sanya Division (Sanya Stomatology Center); Center for Digital Dentistry, Peking University School and Hospital of Stomatology & National Center for Stomatology & National Clinical Research Center for Oral Diseases & National Engineering Research Center of Oral Biomaterials and Digital Medical Devices & Beijing Key Laboratory of Digital Stomatology & NHC Key Laboratory of Digital Stomatology, Beijing 100081, China.
Key points
From the paper's conclusion- Autonomous robotic buccal alveolar bone reduction reproduced the planned crown-lengthening geometry with submillimetre point deviations in this in vitro model, demonstrating the potential of this approach to enhance surgical precision. Further clinical studies are needed to validate its accuracy.
Abstract
OBJECTIVES: To evaluate the planned-to-actual accuracy of autonomous robotic buccal alveolar bone reduction for anterior crown-lengthening surgery in vitro. METHODS: Six standardised maxillary anterior typodont models were digitised using intraoral scanning and cone-beam computed tomography. Buccal reduction volumes were planned for teeth 13-23 (36 tooth-level reduction volumes) and executed by an autonomous robotic system. The planned and actual datasets were compared for apical midpoint deviation (AMD), coronal mesial point deviation (CMD), coronal distal point deviation (CDD), directional components, buccal surface deviation, lingual depth deviation, root surface clearance, and reduction volume deviation. Subgroup and correlation analyses were exploratory. RESULTS: AMD was 0.391 ± 0.243 mm, CMD was 0.598 ± 0.372 mm, and CDD was 0.506 ± 0.292 mm. Planned area coverage was 92.765 ± 4.908%, actual area agreement was 85.974 ± 7.145%, and relative area difference was 8.627 ± 10.891%. The relative volume difference was -5.773 ± 22.531%. Directional deviations differed significantly among the three point locations along the X and Y axes (both P < 0.001), but not along the Z-axis (P = 0.841). The root mean positive distance was 0.558 ± 0.177 mm. Negative root-distance values occurred in 4 of 36 tooth-level reduction volumes from three models, with maximum negative root distances ranging from -0.0106 to -0.6470 mm. CONCLUSIONS: Autonomous robotic buccal alveolar bone reduction reproduced the planned crown-lengthening geometry with submillimetre point deviations in this in vitro model, demonstrating the potential of this approach to enhance surgical precision. Further clinical studies are needed to validate its accuracy. CLINICAL SIGNIFICANCE: In this in vitro study, autonomous robotic buccal alveolar bone reduction for anterior crown lengthening achieved submillimetre accuracy. These findings suggest the potential of this approach to enhance surgical precision.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy of Robotic Computer-Assisted Implant Surgery in Fully Edentulous Patients: A Systematic Review and Meta-Analysis.Journal of dentistry · 2026
- Robotic surgery vs dynamic navigation: does greater accuracy make the difference in implant dentistry?Evidence-based dentistry · 2026
- Comparing the accuracy and active drilling time of robot-assisted and dynamic navigation systems in access cavity preparation: an in vitro study.BMC oral health · 2026
- Clinical Accuracy and Patient Experience Following Robotic-Assisted Full-Arch Implant Rehabilitation: A Case Series.Dentistry journal · 2026
- Accuracy of Minimally Invasive Laminate Veneer Preparations Using an Automated Dental Robotic System and Manual Preparation Technique: An In Vitro Study.The International journal of prosthodontics · 2026
- Static, Dynamic and Robotic CAIS in Healed Partially Edentulous Sites: Accuracy and Chair‐Side TimeClinical Implant Dentistry and Related Research · 2026