Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery

Evaluation of the accuracy of zygomaticomaxillary complex fracture reduction and fixation using a fully digital workflow versus conventional protocol: a randomized clinical trial.

Dina Ayman Fayek, Mohamed Mounir, Sameh Tarek Mekhemer

Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Cairo University, PO Box 11553, Cairo, Egypt; Department of Oral and Maxillofacial Surgery, Faculty of Oral and Dental Medicine, Future University in Egypt, Fifth Settlement, New Cairo, Egypt. Electronic address: dina.fayek@dentistry.cu.edu.eg. · Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Cairo University, PO Box 11553, Cairo, Egypt; Department of Oral and Maxillofacial Surgery, School of Dentistry, New Giza University, Giza, Egypt.

Randomised · n=14

Key points

From the paper's conclusion
  • The results showed that, when comparing the virtual reduction of both groups with the postoperative CT scans, the overall accuracy was markedly higher in the digital group (0.290 ± 0.132 mm) than in the conventional group (1.290 ± 0.525 mm), with a mean difference of -1.000 mm (95% CI -1.309 to -0.691; p < 0.001) and a very large effect size (d = -2.612).

Abstract

The zygomatic bone has multiple points of articulation with the craniofacial skeleton, and accurate intraoperative reduction in various planes is very challenging and entirely dependent on the operator's subjective evaluation. This study aimed to investigate the efficiency of a novel fully digital protocol versus the conventional management strategy for zygomaticomaxillary complex fracture reduction and fixation. Twenty-eight patients with acute zygomatic fractures who met the inclusion criteria were recruited; each group of 14 patients was randomly assigned. The intervention group received a fully digital protocol, represented by customized predictive-hole surgical guides and patient-specific fixation plates, while the control group was managed conventionally. Accuracy of bone reduction was the primary outcome. The results showed that, when comparing the virtual reduction of both groups with the postoperative CT scans, the overall accuracy was markedly higher in the digital group (0.290 ± 0.132 mm) than in the conventional group (1.290 ± 0.525 mm), with a mean difference of -1.000 mm (95% CI -1.309 to -0.691; p < 0.001) and a very large effect size (d = -2.612). Clinically, the digital group showed higher patient satisfaction with aesthetics and a shorter operation duration, while no difference in infraorbital nerve numbness was observed between the two groups. Based on the statistical and clinical findings of this study, a fully digital workflow provided a valuable approach for the management of zygomaticomaxillary complex fractures, optimizing aesthetic and functional rehabilitation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers