Direct chromatographic processing of undiluted human plasma on large-channel monolithic supports at physiological temperature.
Jasmina Simović Medica, Marija Begić Kokan, Tomislav Friganović, Petra Grbčić, Lucija Dončević, Marko Filipović, Dajana Gašo-Sokač, Djuro Josić
General Hospital Pula, 52100 Pula, Croatia. · Sartorius BIA Separations d.o.o., Mirce 21, 5270 Ajdovščina, Slovenia.
Key points
From the paper's conclusion- Here, undiluted, centrifuged fresh-frozen plasma was processed at approximately physiological temperature (~36 °C) on two large-channel (6 μm) polymethacrylate monoliths with distinct surface chemistries: a hydrophilic hydroxyl (OH) monolith used to assess direct plasma passage and a quaternary amine (QA) strong anion-exchange monolith used to demonstrate selective factor VIII (FVIII) retention an
Abstract
At low temperatures, increased plasma viscosity and precipitation of some plasma components make direct chromatographic processing of undiluted plasma technically challenging. Dilution is commonly used to reduce viscosity and ease column loading, whereas contact of plasma with chromatographic surfaces represents a separate concern because surface-mediated activation of coagulation and complement can occur. Such additional sample handling increases process volume and processing time and may contribute to analyte losses. Here, undiluted, centrifuged fresh-frozen plasma was processed at approximately physiological temperature (~36 °C) on two large-channel (6 μm) polymethacrylate monoliths with distinct surface chemistries: a hydrophilic hydroxyl (OH) monolith used to assess direct plasma passage and a quaternary amine (QA) strong anion-exchange monolith used to demonstrate selective factor VIII (FVIII) retention and enrichment. The OH monolith permitted passage of 50 column volumes of undiluted plasma without an observable upward trend in backpressure. In the OH flow-through, FVIII activity (124.3-126.2%; reference range 70-150%), prothrombin time (83-86%; 80-123%), activated partial thromboplastin time (28.6-29.1 s; 21.8-32.6 s), C1-inhibitor concentration (0.25 g/L; 0.15-0.35 g/L), C3 (1.01 g/L; 0.83-1.93 g/L), and C4 (0.26 g/L; 0.15-0.57 g/L) were within laboratory reference ranges, with no systematic changes observed. On the QA monolith, FVIII activity was below the assay detection limit (<5%) in the initial flow-through sample and progressively reappeared at higher cumulative load volumes; in the elution fractions, FVIII-specific activity increased 20.8-35.7-fold. These results provide proof of concept for direct chromatographic processing of undiluted plasma at approximately physiological temperature without prior dilution or precipitation-based pretreatment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Chromatography on monolithic supports for biological nanoparticles and large biomolecules: Recent developments and state-of-the-artJournal of Chromatography Open · 2025
- Label-free microfluidic small-volume blood processing for analysis of platelet activation and neutrophil-platelet aggregates.Lab on a chip · 2026
- Repeatability and reproducibility of chromatographic parameters and resulting figures of merit on polyester C-CP fiber columns for the routine isolation and analysis of HEK293-derived extracellular vesicles.Analytical methods : advancing methods and applications · 2026
- Tunable purification of plasma extracellular vesicles by multimodal exclusion chromatography for bioanalysis.Talanta · 2026
- Whole Blood Holding at 22 °C: Coagulation Factor V and VIII Activity and Plasma Factor Content per Unit.Cureus · 2026
- 357 Investigation to Determine Root Cause for Aggregate formation in Apheresis-Derived Hematopoietic Progenitor Cell Products Prior to Processing and StorageAmerican Journal of Clinical Pathology · 2026