Long-Term Safety and Procedural Performance of an Interatrial Shunt for Heart Failure According to Ejection Fraction.
Josep Rodés-Cabau, Gregg W Stone, Saibal Kar, Stefan Verheye, Ariel Roguin, Jeroen Bax, Antoni Bayes-Genis, Julio Nunez,
Quebec Heart & Lung Institute, Laval University, Quebec City, Canada; Clínic Barcelona, Barcelona, Spain. Electronic address: josep.rodes@criucpq.ulaval.ca. · Icahn School of Medicine at Mount Sinai, New York, New York, USA.
Key points
From the paper's conclusion- Interatrial shunt implantation was mechanically reliable and durable through 2 years. Safety outcomes differed by HF phenotype, with no evidence of excess clinical safety events in HFrEF but increased adverse events in HFpEF. These findings support phenotype-specific evaluation of interatrial shunt therapy. (REducing Lung congestion symptoms using the v-wavE shunt in adVancEd Heart Failure [RELIEVE-HF]; NCT03499236).
Abstract
BACKGROUND: Long-term device durability and phenotype-specific safety outcomes after interatrial shunting require further characterization. OBJECTIVES: This study aimed to evaluate the procedural success, device durability, and heart failure (HF) phenotype-specific safety outcomes of an interatrial shunt in the RELIEVE-HF trial. METHODS: Randomized patients were assigned 1:1 to shunt or sham control with prespecified stratification by left ventricular ejection fraction (LVEF) into heart failure with reduced ejection fraction (HFrEF) (LVEF ≤40%) and heart failure with preserved ejection fraction (HFpEF) (LVEF >40%). Procedural performance, device durability, serious adverse events (SAEs), major adverse cardiovascular or neurological events (MACNE), major bleeding, and renal outcomes were assessed through 2 years. RESULTS: Among 605 enrolled patients (508 randomized; 97 roll-in), shunt implantation was attempted in 370 and was successful in 369 (99.7%). Among implanted patients with follow-up (n = 348), no device- or procedure-related MACNE occurred through 2 years (0%; 95% CI: 0%-0.9%), and shunt patency was 99.4% (95% CI: 97.9%-99.9%). Prespecified LVEF-stratified analyses demonstrated divergent safety profiles. In HFrEF, shunt therapy was not associated with excess all-cause MACNE, major bleeding, or renal decline and was associated with lower SAE burden. In HFpEF, shunt therapy was associated with higher all-cause MACNE, greater SAE burden, increased major bleeding, and greater renal decline. CONCLUSIONS: Interatrial shunt implantation was mechanically reliable and durable through 2 years. Safety outcomes differed by HF phenotype, with no evidence of excess clinical safety events in HFrEF but increased adverse events in HFpEF. These findings support phenotype-specific evaluation of interatrial shunt therapy. (REducing Lung congestion symptoms using the v-wavE shunt in adVancEd Heart Failure [RELIEVE-HF]; NCT03499236).
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Interatrial Shunt Devices in Heart Failure: A Therapy in Search of the Right Patient.Journal of cardiovascular development and disease · 2026
- Overview of the 2025 FDA circulatory systems devices panel advisory committee meeting on the V-Wave Ventura Interatrial Shunt System.Cardiovascular revascularization medicine : including molecular interventions · 2026
- Transcatheter Atrial Shunting in Patients With Heart Failure: A Systematic Review and Meta-Analysis.Cardiology in review · 2026
- Heart Failure With Reduced Ejection Fraction Is Associated With Greater Cardiovascular Risk Following Lumbar Interbody Fusion.Global spine journal · 2026
- Impact of Timing of Atrial Fibrillation Ablation on AF Recurrence and Clinical Outcomes in Patients With HFpEF and HFrEFJournal of Cardiovascular Electrophysiology · 2025
- Intermittent Hypoxic-Hyperoxic Exposure in Patients with Heart Failure - A Pilot Study.ESC heart failure · 2026