Overview of the 2025 FDA circulatory systems devices panel advisory committee meeting on the V-Wave Ventura Interatrial Shunt System.
Ori Waksman, Edward Ha, Asa Phichachop, Ron Waksman
Section of Interventional Cardiology, Medstar Washington Hospital Center, Washington, D.C., United States of America. · Section of Interventional Cardiology, Medstar Washington Hospital Center, Washington, D.C., United States of America. Electronic address: ron.waksman@medstar.net.
Key points
From the paper's conclusion- The Advisory Panel unanimously concluded that, despite procedural safety, the evidence did not demonstrate a favorable benefit-risk profile for the V-Wave device. The deliberations underscore the challenges of relying on subgroup analyses, recurrent-event endpoints, and post-market data to support approval of permanent intracardiac devices in this population.
Abstract
BACKGROUND: Interatrial shunting is a novel strategy to reduce left atrial pressure and heart-failure (HF) morbidity. The Ventura Interatrial Shunt System (V-Wave) was evaluated in the randomized, sham-controlled RELIEVE-HF trial and reviewed by the U.S. Food and Drug Administration (FDA) Circulatory Systems Devices Panel. OBJECTIVES: To summarize the scientific and regulatory deliberations among the sponsor, FDA, and Advisory Panel regarding the safety, effectiveness, and benefit-risk profile V-Wave. METHODS: We reviewed publicly available FDA documents, sponsor presentations, FDA analyses, and Advisory Panel discussions from the December 2025 meeting, with emphasis on trial design, endpoint selection, subgroup interpretation, and regulatory standards. RESULTS: The sponsor emphasized reductions in recurrent HF events among patients with HF with reduced ejection fraction (HFrEF), supported by prespecified recurrent-event analyses and echocardiographic data, while arguing that the hierarchical win-ratio primary endpoint underestimated benefits. In contrast, the FDA highlighted the failure to meet the prespecified primary endpoint in the intention-to-treat population, the fragility of subgroup findings, lack of outcome improvement, and consistent harm in patients with preserved ejection fraction. Advisory Panel members acknowledged acceptable procedural safety but expressed concern regarding the absence of clear, reproducible efficacy, challenges in phenotype-based labeling, and the inability of post-approval registries to resolve evidence gaps. CONCLUSIONS: The Advisory Panel unanimously concluded that, despite procedural safety, the evidence did not demonstrate a favorable benefit-risk profile for the V-Wave device. The deliberations underscore the challenges of relying on subgroup analyses, recurrent-event endpoints, and post-market data to support approval of permanent intracardiac devices in this population.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Long-Term Safety and Procedural Performance of an Interatrial Shunt for Heart Failure According to Ejection Fraction.JACC. Heart failure · 2026
- Interatrial Shunt Devices in Heart Failure: A Therapy in Search of the Right Patient.Journal of cardiovascular development and disease · 2026
- Pulsed Field Ablation for Ventricular Arrhythmias: A Systematic Review and Dual-Framework (Frequentist and Bayesian) Meta-Analysis.Cardiology in review · 2026
- Right heart function paradoxically predicts systemic circulatory failure during ventricular tachyarrhythmias in LVAD-supported patientsEuropean Heart Journal · 2026
- Vericiguat outcomes by pre-randomisation worsening heart failure events: a post-hoc recurrent events analysis of the VICTOR trialEuropean Heart Journal · 2026
- Limited Discrimination of Risk Scores for Right Ventricular Failure After Left Ventricular Assist Device Implantation.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026