Journal of biomechanics

Quantifying digital volume correlation strain uncertainty at the bone-pedicle screw interface: an ex-vivo zero-strain analysis.

Léa Slive, Ye Young Shin, Martine McGregor, Stewart McLachlin, Jonathan Kusins

Department of Systems Design Engineering, University of Waterloo, Waterloo, Ontario, Canada. · Department of Mechanical and Mechatronics Engineering, University of Waterloo, Waterloo, Ontario, Canada.

Key points

From the paper's conclusion
  • Toward improving pre-clinical assessment of pedicle screw fixation, this study combines micro-CT imaging, U-Net deep learning segmentation, and digital volume correlation (DVC) techniques to quantify the errors associated with measuring full-field strains at the screw-bone interface under nominal zero-strain conditions.

Abstract

Pedicle screws are essential for postoperative construct stability in spinal fusion surgeries. However, screw loosening remains a common clinical complication, often linked to excessive loading that induces damaging trabecular bone strains surrounding the pedicle screw. Toward improving pre-clinical assessment of pedicle screw fixation, this study combines micro-CT imaging, U-Net deep learning segmentation, and digital volume correlation (DVC) techniques to quantify the errors associated with measuring full-field strains at the screw-bone interface under nominal zero-strain conditions. Pedicle screws were surgically implanted into six human lumbar vertebrae specimens. Each specimen underwent micro-CT imaging. For three specimens, paired unloaded scans were acquired and analyzed using a standard DVC uncertainty workflow. To mitigate local bone-metal artifacts, bone segmentation was performed using a U-Net model, and segmentation accuracy was quantified using the Sørensen-Dice similarity coefficient. The U-Net model reproduced manual segmentations with high fidelity (average Dice = 0.94 ± 0.01). DVC accuracy and precision depended on the spatial resolution of strain measurement; at a DVC spatial resolution of 1.8 mm, strain estimates around the screw were reliable with low error (accuracy = 387 ± 110 µε; precision = 222 ± 82 µε). This study establishes the feasibility of a micro-CT-based DVC workflow and characterizes the strain measurement errors within trabecular bone surrounding pedicle screws using a zero-strain analysis. This workflow provides a foundation for future studies involving loaded tests, further optimization of segmentation to suppress artifacts, and validation across larger cohorts and implant designs.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts