JACC. Clinical electrophysiology

Conduction System Pacing in Pediatric Patients: Outcomes From a Multicenter International Cohort.

Óscar Cano, Yerlan Turubayev, David Žižek, Shunmuga Sundaram Ponnusamy, Guram Imnadze, Wim Huybrechts, Pugazhendhi Vijayaraman, Manuel Molina-Lerma,

Hospital Universitari i Politècnic La Fe, Valencia, Spain; Instituto de Investigación Sanitaria La Fe, Valencia, Spain; Centro de Investigaciones Biomédicas en RED en Enfermedades Cardiovasculares (CIBERCV), Madrid, Spain. Electronic address: cano_osc@gva.es. · Interventional Arrhythmology Department, "University Medical Center" Corporate Fund, Astana, Kazakhstan.

Observational · n=120

Key points

From the paper's conclusion
  • CSP is a feasible, safe, and durable physiological pacing strategy in children, with excellent long-term CSP preservation and stable electrical performance despite ongoing somatic growth. CSP may also provide effective physiological resynchronization in pediatric patients requiring CRT.

Abstract

BACKGROUND: Conduction system pacing (CSP) may be advantageous in children requiring lifelong ventricular pacing, but long-term durability data are limited. OBJECTIVES: The goal of this study was to evaluate the long-term preservation, durability, and safety of CSP in pediatric patients. METHODS: This international, multicenter, retrospective study included consecutive patients aged ≤18 years undergoing His bundle pacing or left bundle branch area pacing at 9 experienced centers. The primary endpoint was long-term preservation of CSP. RESULTS: Among 120 patients (mean age 12.6 ± 3.7 years), 68 (56.7%) had congenital atrioventricular block, 41 (34.2%) had congenital heart disease, and 12 (10.0%) underwent CSP for heart failure/cardiac resynchronization therapy (CRT). Acute CSP implantation, defined as achievement of His bundle pacing, left bundle branch pacing, or left ventricular septal pacing, was successful in 116 patients (96.7%). After a median follow-up of 24 months (Q1-Q3: 10-54 months), electrical lead performance remained stable. Among 101 patients with CSP at implantation and follow-up capture assessment, CSP was preserved in 95 (94.1%). Kaplan-Meier-estimated freedom from loss of CSP was 100.0% at 1 year and 98.2% (95% CI: 94.7%-100.0%) at 2 and 3 years. CSP lead-related complications occurred in 10 patients (8.3%), and reintervention was required in 9 (7.5%). In the CRT subgroup, left ventricular ejection fraction improved from 28.8% ± 9.4% to 46.8% ± 12.0% (P < 0.001). CONCLUSIONS: CSP is a feasible, safe, and durable physiological pacing strategy in children, with excellent long-term CSP preservation and stable electrical performance despite ongoing somatic growth. CSP may also provide effective physiological resynchronization in pediatric patients requiring CRT.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Conduction System Pacing in Pediatric Patients: Outcomes From a Multicenter International Cohort — Science Skim