JACC. Case reports

Variable Electrogram Detected in the Vein of Marshall After Pulsed Field Application.

Tatsuya Sakamoto, Iwanari Kawamura, Masateru Takigawa, Shinsuke Miyazaki, Tetsuo Sasano

Department of Cardiovascular Medicine, Institute of Science Tokyo, Tokyo, Japan. Electronic address: tatsuya54.sakamoto40@gmail.com. · Department of Cardiovascular Medicine, Institute of Science Tokyo, Tokyo, Japan.

Key points

From the paper's conclusion
  • Reversible effect of pulsed field application on electrograms in the vein of Marshall (VOM).

Abstract

IMAGE: Reversible effect of pulsed field application on electrograms in the vein of Marshall (VOM). CASE SUMMARY: Focal pulsed field ablation at the mitral isthmus attenuated the VOM electrograms. However, the electrogram amplitudes recovered after 4.5 minutes, accompanied by reconduction across the mitral isthmus line. Additional applications eliminated the VOM electrograms, which did not recur thereafter. TAKE-HOME MESSAGES: Pulsed field ablation may be more likely than radiofrequency ablation to cause reversible attenuation of electrogram amplitude, even in epicardial tissue. Even when conduction block is achieved, it may be transient or incomplete; therefore, careful observation and, when necessary, repetitive applications are warranted.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers