Inadvertent Left Ventricular Pacing Through an Unrecognized Sinus Venosus Atrial Septal Defect.
Simon Viscogliosi, Marie Vincent, Géraldine Vedrenne, Raphaël Mirailles, Théo Pezel, Hélène Thibault, Paul-Jun Martial
MIRACL.ai Laboratory, Multimodality Imaging for Research and Artificial Intelligence Core Laboratory, University Hospital of Lariboisiere (AP-HP), Paris, France; Department of Cardiology, University Hospital of Lariboisiere, (Assistance Publique des Hôpitaux de Paris, AP-HP), Université Paris Cité, Paris, France; Inserm MASCOT - UMRS 942, University Hospital of Lariboisiere, Paris, France; Institut de Cardiologie, Lyon University Hospital (Hospices Civils de Lyon), Lyon, France. · Department of Cardiology, University Hospital of Lariboisiere, (Assistance Publique des Hôpitaux de Paris, AP-HP), Université Paris Cité, Paris, France.
Key points
From the paper's conclusion- Seven months later, imaging unexpectedly showed the lead coursing from the superior vena cava through an unrecognized superior sinus venosus defect into the left atrium, across the mitral valve, into the LV.
Abstract
BACKGROUND: Inadvertent left ventricular (LV) lead placement is rare but exposes patients to systemic thromboembolism. CASE SUMMARY: An 89-year-old man received a dual-chamber pacemaker for atrial fibrillation with paroxysmal complete heart block. Severe right atrial dilatation prevented right ventricular lead placement, and the operator converted to a sheath-delivered "left bundle branch area" approach. Seven months later, imaging unexpectedly showed the lead coursing from the superior vena cava through an unrecognized superior sinus venosus defect into the left atrium, across the mitral valve, into the LV. Transesophageal echocardiography confirmed a 30 × 20 mm superior sinus venosus defect with partial anomalous pulmonary venous return. DISCUSSION: A paced right bundle branch block pattern is expected during left bundle branch area pacing, and activation-time criteria may be falsely reassuring when the lead paces the LV endocardium directly. Atrial septal defect screening is warranted whenever right-sided dilatation is unexplained.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Left bundle branch pacing in double-chambered left ventricle and perimembranous ventricular septal aneurysm: a case report.European heart journal. Case reports · 2026
- Case Report: Double-click on the ventricle: ventricular safety pacing triggered by right atrial lead dislodgement in a patient after right atrial tumor resection.Frontiers in cardiovascular medicine · 2026
- Dual-Chamber Leadless Pacemaker Implantation Via a Left-Sided Inferior Vena Cava After Pacemaker Extraction.JACC. Case reports · 2026
- Leadless Pacemaker Implant Location as a Predictor of Pacing Induced Cardiomyopathy.Pacing and clinical electrophysiology : PACE · 2026
- Right-Sided Dual-Chamber Left Bundle Branch Area Pacing in Persistent Left Superior Vena Cava: A Case Report on Personalized Conduction System Pacing.Case reports in cardiology · 2026
- First-in-Human Permanent Totally Leadless Left Ventricular Septal Endocardial Pacing With Atrioventricular Synchrony.JACC. Case reports · 2026