Feasibility and Safety of Transthoracic Echocardiography-Guided Transvenous Permanent Pacemaker Implantation Without Fluoroscopy.

Tianjie Feng, Ziping Li, Jie Dong, Guangzhi Zhao, Cheng Wang, Jing Dong, Xiangbin Pan

Department of Structural Heart Disease, Fuwai Hospital & National Center for Cardiovascular Disease, State Key Laboratory of Cardiovascular Disease, Chinese Academy of Medical Sciences & Peking Union Medical College, National Clinical Research Center for Cardiovascular Diseases, Beijing, China. · Department of Structural Heart Disease, Fuwai Hospital & National Center for Cardiovascular Disease, State Key Laboratory of Cardiovascular Disease, Chinese Academy of Medical Sciences & Peking Union Medical College, National Clinical Research Center for Cardiovascular Diseases, Beijing, China. Electronic address: xiangbin_pan@163.com.

Trial · n=52

Key points

From the paper's conclusion
  • In selected patients with adequate supine acoustic windows, TTE-guided fluoroscopy-free pacemaker implantation was feasible with acceptable acute outcomes. The lead-position findings and exploratory case-order association require further validation in larger and more diverse populations.

Abstract

BACKGROUND: Fluoroscopy is routinely used for transvenous permanent pacemaker implantation, but a standardized transthoracic echocardiography (TTE)-only workflow has not been prospectively evaluated. OBJECTIVES: The objective of the study was to evaluate the feasibility, acute safety, and procedural performance of fluoroscopy-free transvenous pacemaker implantation guided exclusively by TTE. METHODS: In this prospective, single-center feasibility study (RADIANCE [Radiation-free Anatomical Direct Imaging Approach for No-fluoroscopy Cardiac Electrode Implantation]; NCT07458724), 52 consecutive patients referred to the Department of Structural Heart Disease were screened. Thirty-one selected patients (59.6%) with guideline-based pacing indications and adequate prespecified supine TTE acoustic windows underwent attempted TTE-only implantation. Acute procedural success required completion without fluoroscopic conversion and fulfillment of prespecified TTE-defined anatomical, electrical, and safety criteria. RESULTS: TTE-only implantation was successful in 30/31 patients (96.8%; 95% CI: 83.3-99.9), with fluoroscopic conversion in 1/31 (3.2%). The mean procedure time was 55.4 ± 5.6 minutes. No major acute procedural complications occurred. During a median follow-up of 8.8 months (IQR: 6-12 months), atrial lead dislodgement requiring revision occurred in 1/21 dual-chamber recipients (4.8%; 95% CI: 0.1-23.8). Procedure time decreased with chronological case order (-0.59 minutes/case; Newey-West 95% CI: -0.68 to -0.49), an exploratory team-level association. CONCLUSIONS: In selected patients with adequate supine acoustic windows, TTE-guided fluoroscopy-free pacemaker implantation was feasible with acceptable acute outcomes. The lead-position findings and exploratory case-order association require further validation in larger and more diverse populations.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Feasibility and Safety of Transthoracic Echocardiography-Guided Transvenous Permanent Pacemaker Implantation Without Fluoroscopy — Science Skim