Overcoming delivered volume limitations of double-jet impinging atomizers; development of alternative soft mist generation concepts.
Moritz Fleischhauer, Peter Langguth, Bastian Kolb, Luis Marcelo Monrreal Ortega, Felix Weiland, Herbert Wachtel
Boehringer Ingelheim Pharma GmbH & Co. KG, Binger Straße 173, Ingelheim am Rhein, Germany; Pharmaceutical Technology & Biopharmaceutics, University of Mainz, Staudingerweg 5, Mainz, Germany. Electronic address: moritz.fleischhauer.ext@boehringer-ingelheim.com. · Pharmaceutical Technology & Biopharmaceutics, University of Mainz, Staudingerweg 5, Mainz, Germany. Electronic address: langguth@uni-mainz.de.
Key points
From the paper's conclusion- Overall, the single-jet design represents a promising approach to extend the Respimat inhaler family toward lower-potency, less soluble, and shear-sensitive therapeutics.
Abstract
Atomization in the Respimat® relies on impinging microjets. Reorienting the twin jets toward a maximized included angle improves energy transfer, enabling aerosolization at reduced pressure and thereby expanding the design space for higher delivered volumes per actuation. However, this potential optimization is constrained by a trade-off between droplet breakup efficiency and backward-directed spray, causing elevated losses at the nozzle outlet. This trade-off was characterized and motivated the evaluation of alternative soft mist generation concepts. Jet-baffle prototypes were developed in which a single liquid jet impacts perpendicularly (corresponding to 180° between two jets) onto a solid surface. An initial approach with an integrated baffle inside the Uniblock provided high aerosol output but required high internal velocity to generate ∼ 5µm droplets, potentially due to droplet coalescence within the confined near-nozzle region. Repositioning the baffle outside the nozzle provided additional space for atomization, enabled small aerosol droplets and allowed for increased delivered volumes per actuation. Radial droplet dispersion following baffle impact emerged as key driver of mouthpiece deposition and aerosol delivery efficiency. The single-jet concept was translated into functional prototypes. Compared with the standard Respimat device, these prototypes operated at similar jet velocity but lower operating pressure and reduced shear rates within the nozzle microchannel, suggesting gentler aerosolization. Testing with selected biologic formulations indicated better preservation of biologic integrity than with the standard Respimat device and a vibrating mesh nebulizer. Overall, the single-jet design represents a promising approach to extend the Respimat inhaler family toward lower-potency, less soluble, and shear-sensitive therapeutics.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Transitioning from Impinging Jets to Single-Jet Impaction: Investigation of an Alternative Nozzle Design for the Respimat® Soft Mist Inhaler.European journal of pharmaceutical sciences : official journal of the European Federation for Pharmaceutical Sciences · 2026
- Quantification of Fine Spray Evolution in Near Field of Various Swirl Burst Injectors Using High-Speed X-Ray ImagingFire · 2026
- The effect of spray properties and inspiratory flow rate on regional nasal deposition for pressure-swirl versus soft mist devicesInternational Journal of Pharmaceutics · 2026
- Novel swirl core nozzle enhances atomization and suppresses hazardous byproducts of next-generation fire extinguishing agents.Journal of hazardous materials · 2026
- Experimental investigation of air-pressurized atomization characteristics and droplet distribution of liquids with different viscositiesJournal of Physics: Conference Series · 2026
- Microhole wall-profile effects on liquid-bridge dynamics and droplet formation in a vibrating mesh atomizerPhysics of Fluids · 2026