Validating complete-airway computational fluid dynamics (CA-CFD) predictions of regional aerosol deposition in human subjects: comparisons with in vivo data of pharmaceutical aerosol delivery during HFNC therapy.
Arun V Kolanjiyil, Rabijit Dutta, Anya Maradiaga, Casey Grey, Ghali Aladwani, Caleb Dalton, Mohammad A M Momin, Michael Hindle,
Department of Mechanical and Nuclear Engineering, Virginia Commonwealth University, Richmond, VA, United States of America; Department of Pharmaceutics, Virginia Commonwealth University, Richmond, VA, United States of America. · Department of Mechanical and Nuclear Engineering, Virginia Commonwealth University, Richmond, VA, United States of America; Department of Pharmaceutics, Virginia Commonwealth University, Richmond, VA, United States of America. Electronic address: pwlongest@vcu.edu.
Key points
From the paper's conclusion- Through methodical development, a CA-CFD model of aerosol delivery during HFNC therapy was established that closely agrees with the in vivo cohort-level data and can now serve as a "representative population-level digital twin" to optimize aerosol delivery during this highly challenging but clinically popular mode of ventilation support.
Abstract
High-Flow Nasal Cannula (HFNC) is a critical tool for respiratory support in acute respiratory distress syndrome (ARDS) and for subjects with general hypoxemia; however, pharmaceutical aerosol delivery simultaneous with HFNC support is known to be highly inefficient. The aim of this study was to develop and validate a high-fidelity, complete-airway computational fluid dynamics (CA-CFD) model of regional aerosol deposition that occurs when a pharmaceutical aerosol product is administered through a nasal interface simultaneously with HFNC therapy. The CFD model included the aerosol delivery tubing and interface connected to a complete-airway (CA) model extending from the extrathoracic nose-throat (NT) region through the conducting airways and alveoli. CFD-predicted regional aerosol deposition was compared with previously published in vivo experimental data. Predictions of the CA-CFD model agreed closely with the available cohort-level human-subject in vivo data including a predicted aerosol deposition fraction in the lungs of 2.9% (based on nebulized dose) compared with in vivo medians of 2.1% (gamma scintigraphy; range: 1.5-2.9%) and 3.7% (SPECT-CT; range: 2.2-4.7%). In addition to low lung delivery efficiency, the regional drug delivery showed high heterogeneity. Considering a map of regional airway deposition, utilizing a realistic polydisperse droplet size distribution in the stochastic individual path (SIP) models, deposition fractions in the NT, upper tracheobronchial (TB), middle TB, lower TB and alveolar regions were 16%, 0.5%, 0.1%, 0.1%, and 2.4% of the nebulized dose, with ∼ 22% of the nebulized dose exhaled/escaped/dripped, while ∼ 58% was lost in the HFNC circuit. Through methodical development, a CA-CFD model of aerosol delivery during HFNC therapy was established that closely agrees with the in vivo cohort-level data and can now serve as a "representative population-level digital twin" to optimize aerosol delivery during this highly challenging but clinically popular mode of ventilation support.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Transnasal Administration of Surfactant Aerosol to the Lungs of Infants Using Direct Interfaces - Lessons Learned from Computational Fluid Dynamics.Pharmaceutical research · 2026
- Drug Delivery via Vibrating Mesh Nebulizer With Symmetrical and Asymmetrical High Flow Nasal Cannulas in an Adult Model.Respiratory care · 2026
- Modeling exhaled aerosol deposition in an impaction filter across clinically measured breath flowrates.Journal of breath research · 2026
- Impact of gas carrier on inhaled medicinal aerosolized particles transport and deposition in a human respiratory system: A computational fluid particle dynamics studyPhysics of Fluids · 2026
- Experimental and computational study of pMDI aerosol delivery in mucus-coated pediatric airways: Effects of inhalation flow and actuation timing.Journal of pharmaceutical sciences · 2026
- Nebulized Aerosol Characteristics During Nasal High-Flow: Effect of Flow, Formulation and Impactor Temperature.Pharmaceutics · 2026