International journal of medical microbiology : IJMM

Molecular detection of highly virulent Legionella pneumophila: Validation of a novel lag-1 targeted qPCR.

Sarah Uhle, Laurine Wehmeier, Susan Wünsche, Anastasia Flountzi, Sophie Jarraud, Markus Petzold, ESCMID Study Group for Legionella Infections (ESGLI)

Institute for Medical Microbiology and Virology, Medical Faculty and University Hospital Carl Gustav Carus, TUD Dresden University of Technology, Fetscherstr. 74, Dresden 01309, Germany; European Study group on Legionella Infections (ESGLI) of the European Society of Clinical Microbiology and Infectious Disease (ESCMID), Aeschenvorstadt 55, Basel 4051, Switzerland; European Union Reference Laboratory for public health in the field of Legionella (EURL-PH-LEGI), Coordinated in Lyon, Hospices Civils de Lyon, 103 Grande rue de la Croix Rousse, Lyon 69317, France. Electronic address: Sarah.uhle@ukdd.de. · Institute for Medical Microbiology and Virology, Medical Faculty and University Hospital Carl Gustav Carus, TUD Dresden University of Technology, Fetscherstr. 74, Dresden 01309, Germany; European Study group on Legionella Infections (ESGLI) of the European Society of Clinical Microbiology and Infectious Disease (ESCMID), Aeschenvorstadt 55, Basel 4051, Switzerland; European Union Reference Laboratory for public health in the field of Legionella (EURL-PH-LEGI), Coordinated in Lyon, Hospices Civils de Lyon, 103 Grande rue de la Croix Rousse, Lyon 69317, France. Electronic address: Laurine.wehmeier@ukdd.de.

Key points

From the paper's conclusion
  • The qPCR assay demonstrated complete concordance and amplified all lag-1 variants without cross-reactivity during validation. The LoD was 10 GU/reaction, and the LoQ was 50 GU/reaction. Using clinical samples, the assay showed a sensitivity of 95.48% and a specificity of 98.58% compared with mAb-based typing. From 248 specimens, 239 showed a complete concordance between the lag-1 qPCR and mAb-typing CONCLUSION: The lag-1 qPCR assay offers a culture-independent, rapid, and sensitive method for detecting virulent mAb 3/1+L. pneumophila strains. This tool has the potential to enhance diagnostic workflows and guide public health interventions by integrating virulence markers into risk assessment protocols.

Abstract

BACKGROUND: Legionella pneumophila is the causative agent of legionellosis and is primarily associated with L. pneumophila serogroup 1 (Lp1). Strains reactive with the monoclonal antibody (mAb) 3/1 are associated with higher virulence and are predominantly responsible for clinical cases. The reactivity with mAb 3/1 correlates with the presence of the lag-1 gene, encoding an O-acetyltransferase that modifies lipopolysaccharides on the bacterial membrane. Despite its clinical importance, rapid and reliable tools for detecting lag-1+ strains are limited. OBJECTIVES: We aimed to develop and validate a quantitative real-time PCR (qPCR) assay targeting the lag-1 gene for detecting mAb 3/1+L. pneumophila strains. METHODS: Primers and probes were designed to amplify all known lag-1 allelic variants. The assay was validated with regard to specificity, sensitivity, limit of detection (LoD), and quantification (LoQ). Clinical evaluation was performed on 249 respiratory samples, comparing lag-1 qPCR results with mAb-based typing results, especially reactivity with mAb 3/1, of isolates and whole genome sequencing. RESULTS: The qPCR assay demonstrated complete concordance and amplified all lag-1 variants without cross-reactivity during validation. The LoD was 10 GU/reaction, and the LoQ was 50 GU/reaction. Using clinical samples, the assay showed a sensitivity of 95.48% and a specificity of 98.58% compared with mAb-based typing. From 248 specimens, 239 showed a complete concordance between the lag-1 qPCR and mAb-typing CONCLUSION: The lag-1 qPCR assay offers a culture-independent, rapid, and sensitive method for detecting virulent mAb 3/1+L. pneumophila strains. This tool has the potential to enhance diagnostic workflows and guide public health interventions by integrating virulence markers into risk assessment protocols.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers