International journal of cardiology

Long-term safety and efficacy of arrhythmia monitoring with an implantable cardiac monitor.

Eimo Martens, Romain Eschalier, Pilar Cabanas, Laurence Jesel, Claudia Amellone, Paul Milliez, Dennis Lau, Andreas Hain,

Klinikum rechts der Isar, München, Germany. Electronic address: eimo.martens@mri.tum.de. · Centre Hospitalier Universitaire Clermort-Ferrand, Clermont Ferrand, France.

Key points

From the paper's conclusion
  • ICMs continue to deliver clinically meaningful arrhythmia diagnoses in the third and fourth year post-implantation, even when the preceding follow-up has been uneventful. Long-term follow-up and explantation have a low risk. Our results affirm the safety and efficacy over the device's lifetime in clinical practice.

Abstract

BACKGROUND: Implantable cardiac monitors (ICMs) provide long-term monitoring for cardiac arrhythmias. This analysis summarizes late diagnoses, patient management and complications until 5 years after implantation. METHODS: The BIO|STREAM-ICM study is an ongoing, prospective registry capturing the real-world application of ICMs at 38 sites across 9 countries. Patients with any ICM indication are followed until device information is no longer clinically utilized. All arrhythmia diagnoses were assessed for relevance to the device indication. RESULTS: Of 1249 patients included in the present analysis, most received an ICM after syncope (56%) or cryptogenic stroke (30%). Follow-up extended to 5 years, 34% of patients had at least 3 years of monitoring. Kaplan-Meier estimates of the cumulative incidence of indication-related diagnosis were 31%, 36%, and 39% for syncope, and 21%, 24%, and 28% for cryptogenic stroke at 2, 3, and 4 years, respectively. After diagnosis, ICMs were explanted in 46% of cases, continued for monitoring of therapy efficacy in 41%, and inactivated but left in situ in 12%. Explantations needed a median duration of 10 min and were characterized as easy or very easy in 86% of cases. Six device related adverse events occurred later than 180 days after insertion, and none in the context of explantations. CONCLUSION: ICMs continue to deliver clinically meaningful arrhythmia diagnoses in the third and fourth year post-implantation, even when the preceding follow-up has been uneventful. Long-term follow-up and explantation have a low risk. Our results affirm the safety and efficacy over the device's lifetime in clinical practice.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Long-term safety and efficacy of arrhythmia monitoring with an implantable cardiac monitor — Science Skim