Precise anatomical landmark annotation of the cardiac conduction system on cardiac computed tomography angiography.
Justin T Tretter, Tarikh Asyraf, Maialen Ugarteburu, Stefano Spaziani, Nicole Perez-Salas, Nathan Webb, Lev Zaretsky, Shlomo Ben-Haim
Hobart Healthcare Research Institute, London, United Kingdom. Electronic address: trettej3@ccf.org. · CARA Medical Ltd., Rochdale, United Kingdom.
Key points
From the paper's conclusion- The method meets its clinically accepted margins with millimeter precision and good-to-excellent reliability, and transfers from expert to trained readers with bias within margin.
Abstract
BACKGROUND: CTA-based estimation of the conduction system from anatomical landmarks underpins a visualization platform (CARA) used in structural heart interventions and conduction system pacing. OBJECTIVE: To quantify the precision and reliability of the annotation method and the bias of trained readers relative to an expert reference standard. METHODS: The expert who defined the method and 6 trained readers each annotated 8 landmarks representing the sinus and atrioventricular nodes, His bundle course (Point B), left bundle branch origin (Point C), and course of Bachmann's bundle (B2-B5), in addition to the three aortic leaflet nadirs on 50 CTA studies in 3 sessions. Repeatability and reproducibility were assessed, referenced to clinically accepted error margins (2-5 mm). Intraclass correlations (ICC) were computed on heights above the aortic annular plane. 3D bias relative to the expert was tested for equivalence within half the margin. RESULTS: Repeatability SD was 1.67-3.98 mm and reproducibility SD 2.03-5.60 mm; all mean deviations were within margin, with 88.6-97.4% of repeat reads and 71.4-94.9% of reader means within margin. Inter-observer ICC on annular heights was 0.76-0.82 for the atrioventricular and 0.79-0.98 for the atrial points. Shared bias relative to the expert was 0.14-3.24 mm, within margin for all landmarks and equivalent within half the margin for sinus node, Point B, B2 and B3; at B3 and B2, 95% and 83% of the apparent separation was repeatability noise. CONCLUSION: The method meets its clinically accepted margins with millimeter precision and good-to-excellent reliability, and transfers from expert to trained readers with bias within margin.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Patient-specific Visualisation of the Cardiac Conduction System to Guide Structural Interventions and Physiological Pacing.Arrhythmia & electrophysiology review · 2026
- Accuracy of CT-derived interarterial and intramural anomalous coronary arteries anatomy compared with 3D-reconstructed IVUS in-vivo reference.The international journal of cardiovascular imaging · 2026
- Accuracy of CT-derived anomalous coronary arteries anatomy versus 3D-reconstructed IVUS in-vivo referenceEuropean Heart Journal · 2026
- Quantitative CT Lead Localization Reveals Proximal Left Bundle Targeting Facilitates Superior Physiologic Pacing.Heart rhythm · 2026
- The anatomy of atrial conduction: A review of anatomic landmarks integrated with computed tomographic virtual dissection to provide a road map for right atrial pacingHeart Rhythm · 2025
- Segment-Level Evaluability of Dedicated Non-ECG-Gated Coronary CT Angiography: Associations with Calcification, Heart Rate, Vessel Territory, and Caliber.Diagnostics (Basel, Switzerland) · 2026