Accuracy of CT-derived anomalous coronary arteries anatomy versus 3D-reconstructed IVUS in-vivo reference
A W Stark, A Mokhtari, R Kakizaki, M R Bigler, M Ilic, A Ashraf, K Stoyanov, C Ryffel,
Key points
From the paper's conclusion- Conclusions When using 3D-IVUS serving as an in vivo reference standard for precise coronary anatomy assessment in AAOCA, shows that CCTA reliably quantifies [GC1] ostial dimensions and IML using the ILS method especially for larger vessel dimensions, supporting its role as a non-invasive classification tool, but remains unreliable for MLA position-dependent measurements and submillimeter precisio
Abstract
Background In anomalous aortic origin of a coronary artery (AAOCA), coronary CT angiography (CCTA) is the primary tool for anatomical assessment. However, CCTA measurements lack validation against a high-resolution in-vivo standard. Objectives To assess CCTA reliability for lumen dimensions and intramural length (IML) in AAOCA, and to validate these against a three-dimensionally reconstructed intravascular ultrasound (3D-IVUS) reference standard using additionally three IML methodologies. Methods Fifty consecutive AAOCA patients underwent both CCTA and invasive IVUS. A 3D-IVUS reference standard was constructed and its geometric reproducibility assessed. CCTA reliability and agreement with 3D-IVUS were evaluated for ostial dimensions, position-dependent minimal lumen area (MLA), and IML, measured by interluminal space (ILS), elliptic ratio, and pericoronary fat sign. Results 3D-IVUS reference standard showed excellent geometric reproducibility (median point-wise distance 0.21 mm, 95%CI 0.08–0.44 mm) and high intrareader reliability (ICC 0.91–0.97). CCTA intra- and inter-reader reliability was good-to-excellent for ostial parameters and ILS-derived IML (ICC 0.83–0.93), but poor for MLA position (ICC 0.32–0.53). Against 3D-IVUS, CCTA ostial dimensions correlated strongly (r=0.71, p<0.001) with minimal bias, and ILS-derived IML showed acceptable agreement (r=0.64; bias +4.8%), while the fat sign substantially overestimated length (+94.7%). Conclusions When using 3D-IVUS serving as an in vivo reference standard for precise coronary anatomy assessment in AAOCA, shows that CCTA reliably quantifies [GC1] ostial dimensions and IML using the ILS method especially for larger vessel dimensions, supporting its role as a non-invasive classification tool, but remains unreliable for MLA position-dependent measurements and submillimeter precision measurements.For image description, please refer to the figure legend and surrounding text.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy of CT-derived interarterial and intramural anomalous coronary arteries anatomy compared with 3D-reconstructed IVUS in-vivo reference.The international journal of cardiovascular imaging · 2026
- Evaluation of High-Risk Anatomical Features in Single and Opposite-Sinus Origin Coronary Artery Anomalies Using Coronary Computed Tomography Angiography.Diagnostics (Basel, Switzerland) · 2026
- Geometric shape-derived CCTA resistance metrics predict invasive FFR in anomalous right coronary arteries under adenosine and dobutamine stressEuropean Heart Journal · 2026
- Predictive value of CT-FFR in right anomalous aortic origin of coronary artery in an adult populationEuropean Heart Journal · 2026
- Imaging-based identification of high-risk anomalous aortic origin of coronary arteries (AAOCA) patientsEuropean Heart Journal · 2026
- Opportunistic volume assessment of left ventricular volume from coronary CT angiography: agreement with cardiac MRIEuropean Heart Journal · 2026