Patient-Reported Symptoms and Arrhythmias During Continuous ECG Monitoring After Transcatheter Aortic Valve Implantation.
Arianna Sofia Bettelini, Céline Moser, Jonas Brügger, Fabian Jordan, Ida Nussbaumer, Leon Scharf, Sven Knecht, Teodor Serban,
Department of Cardiology, University Hospital Basel, Basel, Switzerland; Cardiovascular Research Institute Basel, University Hospital Basel, Basel, Switzerland. · Department of Cardiology, University Hospital Basel, Basel, Switzerland.
Key points
From the paper's conclusion- After TAVI, patient-triggered events during ECG patch monitoring poorly reflect objectively documented arrhythmias and fail to identify many clinically relevant rhythm disturbances. These findings support systematic continuous ECG surveillance beyond symptom-driven monitoring strategies after TAVI.
Abstract
BACKGROUND: Wearable ECG patches after transcatheter aortic valve implantation (TAVI) enable continuous rhythm monitoring and allow symptoms reporting via trigger buttons and diary entries. However, the reliability of triggered events to identify clinically relevant arrhythmias after TAVI is uncertain. OBJECTIVES: We systematically analyzed patient-triggered events and their corresponding rhythm findings during 14-day ECG patch monitoring after TAVI. METHODS: Adult patients undergoing TAVI at a Swiss tertiary centre were prospectively enrolled and monitored with a 14-day ECG patch before discharge. Patients recorded any cardiac-related symptoms using the patch trigger button and/or diary function. RESULTS: Among 315 patients (median age 82 years [78-87], 43% female), median analyzable duration was 14 days [13-14]. At least one patient-triggered event was recorded in 94 patients (30%), with a median of 2 events per patient [1-4]. Atrial fibrillation (AF) was detected in 85 patients (27%), including 17 patients (5.4%) with new AF; however, only 20 patients with AF (24%) reported symptoms during AF episodes. Pauses occurred in 58 patients (18%) and high-grade atrioventricular block in 32 patients (10%), with no associated patient-triggered events. Ventricular tachycardia (≥4 beats) occurred in 165 patients (52%) and was entirely asymptomatic. Of 348 evaluable symptom-triggered events, 153 (44%) occurred without associated arrhythmia. The most frequent arrhythmias associated with symptom-triggered events were premature atrial contractions (21%), AF (14%), and premature ventricular contractions (11%). CONCLUSIONS: After TAVI, patient-triggered events during ECG patch monitoring poorly reflect objectively documented arrhythmias and fail to identify many clinically relevant rhythm disturbances. These findings support systematic continuous ECG surveillance beyond symptom-driven monitoring strategies after TAVI.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Diagnostic yield of patient-triggered events during 14-days-electrocardiographic monitoring after transcatheter aortic valve implantationEuropean Heart Journal · 2026
- Time to new-onset atrial fibrillation after transaortic valve implantation using 14-days ambulatory electrocardiographic monitoringEuropean Heart Journal · 2026
- Paired 14-day ECG monitoring reveals divergent arrhythmia dynamics before and after transcatheter aortic valve implantationEuropean Heart Journal · 2026
- Extended Wearable ECG Monitoring Detects Unrecognised Arrhythmias After COPD Exacerbation: The PRECURSOR Pilot Study.The clinical respiratory journal · 2026
- One-week patch ECG reveals residual atrial fibrillation missed by repeated Holter monitoring after apparently successful catheter ablation.Journal of electrocardiology · 2026
- Silent conduction disturbances after TAVI: insights from 12-lead holter monitoringEuropean Heart Journal · 2026