Ten-year Performance of the Multi-Layer Insulated Defibrillation Lead in the Medicare Population.
Robert D Schaller, Haris Haqqani, Ulrika Birgersdotter-Green, Thomas Callahan, Amy Jan, Wenwen Li, Yelena Nabutovsky, Vish Charan,
Section of Cardiac Electrophysiology, Cardiovascular Division, Department of Medicine, Hospital of the University of Pennsylvania, Philadelphia, PA. Electronic address: robert.schaller@pennmedicine.upenn.edu. · Department of Cardiology, The Prince Charles Hospital, Brisbane, Australia; Faculty of Medicine, University of Queensland, Brisbane, Australia.
Key points
From the paper's conclusion- The Optisure ICD lead demonstrated favorable mechanical performance through 10 years in routine clinical practice. These findings support both the long-term durability of Optisure and the use of integrated real-world data for post-market device surveillance.
Abstract
BACKGROUND: Long-term surveillance of implantable cardioverter-defibrillator (ICD) leads is essential because clinically important mechanical failures may emerge years after implantation. OBJECTIVE: To evaluate 10-year mechanical complication-free survival of the Optisure ICD lead (Abbott, Sylmar, CA) in a large real-world Medicare population. METHODS: Manufacturer device-tracking records were linked to Medicare Fee-for-Service claims and device complaint data using the EP PASSION framework. Patients receiving an Optisure lead from February 2014 through December 2024 were included. Mechanical lead-related complications were defined by claims containing both a diagnosis of mechanical lead malfunction and a lead-related surgical intervention more than 30 days after implantation. Kaplan-Meier survival was estimated under a best-case scenario including confirmed Optisure-related events and a worst-case scenario also including potentially related events. RESULTS: The cohort included 26,830 patients with 27,033 Optisure leads. Mean age was 73.8 ± 8.6 years, 28.4% were women, and 49.8% had cardiac resynchronization therapy ICDs. Overall, 379 complications were classified as confirmed lead-related and 606 as confirmed or potentially lead-related. At 10 years, complication-free lead survival was 97.2% (95% confidence interval [CI], 96.7%-97.7%) in the best-case scenario and 95.6% (95% CI, 94.9%-96.2%) in the worst-case scenario. Survival did not differ significantly by age, lead length, coil configuration, or device type. CONCLUSION: The Optisure ICD lead demonstrated favorable mechanical performance through 10 years in routine clinical practice. These findings support both the long-term durability of Optisure and the use of integrated real-world data for post-market device surveillance.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Safety and Efficacy of the Novel OmniaSecure Defibrillation Lead Through Long-Term Follow-Up: Final Results From the LEADR TrialCirculation Arrhythmia and Electrophysiology · 2025
- Clinical outcomes of transvenous lead extraction in octogenarians: a real-world propensity score-matched analysisEuropean Heart Journal · 2026
- Evaluation of primary prevention ICD efficacy in late elderly heart failure patients: a multicenter studyEuropean Heart Journal · 2026
- Age and triggers of inappropriate implantable cardioverter-defibrillator therapy in primary prevention patientsEuropean Heart Journal · 2026
- Eighteen Years of Complication-Free Survival Following Inadvertent Left Ventricular Implantable Cardioverter Defibrillator (ICD) Lead Placement via Atrial Septal Defect: A Case Report.Cureus · 2026
- Real-world outcomes of subcutaneous ICD implants: a single-centre experienceEuropean Heart Journal · 2026