The Complex Relationship Between Volume and Outcome of AMS 800 Implants in the SATURN Database.
Frank Van der Aa, Nikesh Thiruchelvam, John Heesakkers, Rizwan Hamid, Frank Martens, Koenraad Van Renterghem, Riccardo Bientinesi, Roman Zachoval,
Department of Urology, University Hospitals Leuven (UZ Leuven), Leuven, Belgium. Electronic address: frank.vanderaa@uzleuven.be. · Department of Urology, Addenbrooke's Hospital, Cambridge University Hospitals NHS Foundation Trust, Cambridge, UK.
Key points
From the paper's conclusion- Although crude analyses showed differences in complication and revision rates across volume groups, the complication pattern was nonmonotonic, with the intermediate-volume group representing the main outlier, whereas revision-related outcomes were higher in both intermediate- and high-volume centers than in low-volume centers. After multivariable adjustment and accounting for center-level clustering, no evidence of an independent center-volume effect was observed for any of the four 1-yr outcomes.
Abstract
BACKGROUND: SATURN is a prospective, multicenter registry that collects data from 28 European centers. Contrary to expectations, previous analyses of the SATURN registry showed higher revision rates in higher-volume centers. OBJECTIVE: We further evaluated the relationship between center volume and 1-yr safety outcomes after AMS 800 implantation using real-world SATURN data. DESIGN, SETTING, AND PARTICIPANTS: We reviewed 1046 SATURN implantations and included 684 patients who received an AMS 800 artificial urinary sphincter from 25 European centers. Centers were categorized by annual AMS 800 implant numbers as low (<10), intermediate (10-25), or high (>25). OUTCOME MEASUREMENTS AND STATISTICAL ANALYSIS: The 1-yr complication, revision, and explant rates were assessed using Kaplan-Meier and log-rank tests and Pearson chi-square tests, with multivariable Cox regression adjusted for relevant patient- and procedure-related covariates and accounting for center-level clustering. RESULTS AND LIMITATIONS: Crude analyses differed by center volume for complications, revisions other than explantation, and all revisions, including explantation, but not for explants. Complication rates were 16.7%, 25.8%, and 17.8% in low-, intermediate-, and high-volume centers, respectively, demonstrating a nonmonotonic pattern with the intermediate-volume group as the main outlier. Rates of revision other than explantation were 5.1%, 12.3%, and 11.2%, respectively, whereas rates of all revisions including explantation were 9.6%, 18.9%, and 16.9%, respectively, and were thus higher in both intermediate- and high-volume centers than in low-volume centers. After adjustment, there was no statistically significant global association between center volume and any of the four 1-yr outcomes; the observed crude differences were not retained after adjustment. This observational registry analysis is subject to confounding and cannot establish causality. CONCLUSIONS: Although crude analyses showed differences in complication and revision rates across volume groups, the complication pattern was nonmonotonic, with the intermediate-volume group representing the main outlier, whereas revision-related outcomes were higher in both intermediate- and high-volume centers than in low-volume centers. After multivariable adjustment and accounting for center-level clustering, no evidence of an independent center-volume effect was observed for any of the four 1-yr outcomes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- One-year Outcomes of Female Artificial Urinary Sphincter Implantation in Europe: The VENUS Registry.European urology focus · 2026
- Temporal trends in patients undergoing artificial urinary sphincter implantation: explantation outcomes and early failure patterns in a tertiary reconstructive referral center.World journal of urology · 2026
- Day-case Artificial Urinary Sphincter Implantation in a Low-Volume Centre: Feasibility With an Experienced Surgeon.The French journal of urology · 2026
- The Artificial Urinary Sphincter Clinical Outcomes Trial: Primary ResultsThe Journal of Urology · 2025
- Volume-outcome analysis of in-hospital complications after cardiac implantable electronic device implantation: a 250 000-case study‡.European heart journal. Quality of care & clinical outcomes · 2026
- Artificial Urinary Sphincter in Male Patients Requiring Urethral Instrumentation: Is It a Challenge?Neurourology and urodynamics · 2026