Endocrinologia, diabetes y nutricion

Diabetic ketoacidosis associated with the reuse of single-use materials in a patient with type 1 diabetes mellitus treated with an advanced hybrid closed-loop system: A case report.

Alba López Calles, Paula Plaza Chaparro, Jesús Moreno Fernández

Servicio de Cardiología, Hospital General Universitario de Ciudad Real, Ciudad Real, Spain. · Medicina Familiar y Comunitaria, Hospital General Universitario de Ciudad Real, Ciudad Real, Spain. Electronic address: pplazac@sescam.jccm.es.

Case report

Key points

From the paper's conclusion
  • We describe the case of a 45-year-old woman using an advanced hybrid closed-loop system who was admitted with an episode of mild DKA in the absence of identifiable precipitating factors.

Abstract

Diabetic ketoacidosis (DKA) remains a potentially serious complication in individuals with type 1 diabetes mellitus, even in those treated with continuous subcutaneous insulin infusion systems. We describe the case of a 45-year-old woman using an advanced hybrid closed-loop system who was admitted with an episode of mild DKA in the absence of identifiable precipitating factors. Metabolic correction was performed according to current recommendations, with a favorable clinical course. Subsequent review of continuous glucose monitoring data and insulin delivery records revealed persistent hyperglycemia secondary to malfunction of the infusion device. This was associated with structural deterioration of a reused insulin cartridge designed for single use, in the context of a shortage of original supplies. This case highlights the importance of infusion materials in pump therapy safety, the limitations of automated systems in detecting mechanical failures, and the need for safe alternative strategies in situations of supply disruption.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers