Toward Trustworthy Diabetes Technologies: A Dynamic Platform for Safety and Security Evaluation.
Anna R Bird, Vivek Mohan, Kexin Wei, Zong Cao, Shi Hoe Ng, Manasvi M Mulay, Viktor Schlegel, Miro G Moffett,
Imperial Global Singapore, Imperial College London, London, UK. · School of Electrical and Electronic Engineering, Nanyang Technological University, Singapore, Singapore.
Key points
From the paper's conclusion- Conclusion: The platform enables system-level cybersecurity assessment of AID under dynamic operating conditions and highlights safety-security interdependencies in closed-loop medical therapeutics.
Abstract
Objective: To evaluate the cybersecurity of automated insulin delivery (AID) systems, a platform was developed, integrating continuous glucose monitors, fluidic systems, insulin controllers, and wireless communication pathways with virtual subjects. Impact Statement: This study provides evidence that AID systems that are vulnerable to cyberattacks pose substantial risk to patient safety. This underscores the need for improved regulation and design of personal, wearable medical devices, through cyberphysiological system modeling. Introduction: AID systems depend on continuous glucose monitors (CGMs) as their primary sensing input, making evaluation of the accuracy, robustness, and security of these sensors essential for safe diabetes management. However, there are limited assessment tools that preserve the communication and timing paths of CGM sensors within an AID system. Methods: We present a modular hardware-in-the-loop platform that reproduces physiological glucose dynamics, bridging the gap between simulation and clinical validation. The system integrates programmable peristaltic pumps, 3D-printed flow cells, glucose sensors, and a real-time virtual subject simulator. Results: The platform generates reproducible glucose trajectories and supports closed-loop operation through integration with an open-source insulin controller. A false data injection attack scenario on the glucose sensor led to an increase in insulin dosage delivered by the control algorithm relative to baseline. If undetected, this could result in a life-threatening condition. Conclusion: The platform enables system-level cybersecurity assessment of AID under dynamic operating conditions and highlights safety-security interdependencies in closed-loop medical therapeutics.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Wearable, Dual Closed‐loop Insulin Delivery System for Precision Diabetes ManagementAdvanced Materials · 2026
- On-device reinforcement learning-based insulin dosing controller for fully automated insulin deliveryFrontiers in Digital Health · 2026
- Artificial Pancreas and Closed-Loop Insulin Delivery: From Early Concepts to AI-Driven Diabetes Automation.Biosensors · 2026
- Patient-Specific Spatio-Temporal False Data Injection Attack Detection for IoMT Using a Graph-GRU Digital Twin and Kalman Innovation Features.Bioengineering (Basel, Switzerland) · 2026
- Current status and future outlook of artificial intelligence and automation in intravenous infusion systems.NPJ digital medicine · 2026
- AI-Enhanced Wearable Glucose Sensors: From Intelligent Sensing to Closed-Loop Management of Diabetes.ACS sensors · 2026