Assessment of beam-control time delay in surface-guided radiotherapy gating using ExacTrac dynamic: impact of surface color and tracking area shape.

Hiroki Katayama, Yosuke Takahashi, Motonori Kitaoka, Hiroki Kawasaki, Takashi Tanii, Marimo Katami, Masato Tsuzuki

Division of Clinical Radiology, Department of Medical Technology, Kagawa University Hospital, Kagawa, Japan. Electronic address: katayama.hiroki@kagawa-u.ac.jp. · Division of Clinical Radiology, Department of Medical Technology, Kagawa University Hospital, Kagawa, Japan.

Key points

From the paper's conclusion
  • For brown surfaces, potential delay increases associated with reduced reflection and impaired surface recognition should be considered during gated radiotherapy. Conversely, the influence of STA shape was minimal, suggesting that STA can be selected based on clinical need. These findings support optimizing gated irradiation using ETD.

Abstract

PURPOSE: Surface-guided radiotherapy (SGRT) enables gated irradiation by synchronizing beam delivery with real-time surface tracking. This study aimed to evaluate the beam-control time delay of the ExacTrac Dynamic (ETD) system. METHODS: Beam-on and beam-off time delays were assessed using a phantom with an embedded metal ball placed at its isocenter. Gated irradiation was performed based on the ETD gating signal. Cine electronic portal imaging device (EPID) images were analyzed to determine the delays. Three phantom surface colors (white, beige, and brown), four skin-tone settings (light, fair, medium, and dark), and three surface tracking area (STA) shapes (small, large, separate) were examined. RESULTS: The beam-on delay was consistently longer than the beam-off delay. The overall mean values of the beam-on and beam-off time delays were 346 ms and 183 ms for the white surface and 353 ms and 181 ms for the beige surface, respectively. No significant differences were observed among the skin-tone settings for either the white or beige phantom surfaces (p > 0.05). For brown surfaces, STA assignment was only possible for the "dark" setting; the beam-on time delay increased to 479 ms, with some values exceeding 500 ms. No significant differences were observed among the three STA shapes (p > 0.05). CONCLUSION: For brown surfaces, potential delay increases associated with reduced reflection and impaired surface recognition should be considered during gated radiotherapy. Conversely, the influence of STA shape was minimal, suggesting that STA can be selected based on clinical need. These findings support optimizing gated irradiation using ETD.