Color K-edge imaging using spectral photon-counting CT with high-Z elements for lung functional imaging: A phantom study.
Agnieszka J Gutwinska, Derick Rosario, Antoine Robert, Eloise Thomas, Amanda Pang, Christian A Hernandez, Ramata S Coulibaly, Klaus Erhard,
Université de Lyon, INSA-Lyon, Université Claude Bernard Lyon 1, UJM-Saint Etienne, CNRS, Inserm, CREATIS UMR 5220, U1206, F-69621, 7 Avenue Jean Capelle O, 69100 Villeurbanne, France. · University of Pennsylvania, Department of Radiology, Philadelphia, United States of America.
Key points
From the paper's conclusion- Color K-edge imaging is feasible with multiple high-Z elements, whilst outperforming the performance of conventional imaging, paving the way for lung SPCCT functional imaging.
Abstract
PURPOSE: To assess the feasibility and performance of color K-edge imaging using spectral photon-counting CT (SPCCT) combined with eight high-Z-elements candidates for lung functional imaging. MATERIAL AND METHODS: Solutions of gadolinium, holmium, ytterbium, hafnium, tantalum, tungsten, gold, and bismuth, were prepared in 12 tubes from 0 to 2 mg/mL, and inserted into an anthropomorphic lung phantom (QRM GmbH). Scans were performed using lung protocols, on a clinical SPCCT (Philips, Israel), at 120 kVp and 3 radiation exposure levels (50, 75, 150 mAs). Noise and contrast-to-noise ratios (CNR) were measured in all images as well as relative percentage errors (RPE). Linear regressions (slope, intercept) of measured versus prepared concentrations and CNR versus concentration were plotted.. RESULTS: On average, mean absolute RPE ranged from 15.9%±8.0% for ytterbium to 34.7%±19.0% for bismuth. Quantification accuracy was excellent for all elements but lower for bismuth. For all elements and exposure levels, mean noise in color K-edge and conventional images was 0.030 ± 0.007 mg/mL and 20.4 ± 2.7 HU, respectively. CNR in color K-edge images increased as the concentrations increased, with an average slope (a) of 43.2 ± 11.8 mL/mg, the highest being for ytterbium (60.1 ± 8.1 mL/mg) and the lowest for bismuth (23.8 ± 5.2 mL/mg). CNR in conventional images also increased as the concentrations increased, with an average slope of 1.7 ± 0.5 mL/mg; the highest was for gadolinium (2.3 ± 0.3 mL/mg) and the lowest for tungsten (1.2 ± 0.1 mL/mg).. CONCLUSION: Color K-edge imaging is feasible with multiple high-Z elements, whilst outperforming the performance of conventional imaging, paving the way for lung SPCCT functional imaging.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- First evaluation of color K-edge image quality using spectral photon-counting CT combined with two contrast agents: A phantom studyDiagnostic and Interventional Imaging · 2025
- K‐Edge imaging using a clinical dual‐source photon‐counting CT systemJournal of Applied Clinical Medical Physics · 2026
- Simultaneous pulmonary ventilation and perfusion imaging using spectral photon counting CT with xenon and gadolinium-based agents in combination with color K-edge imagingDiagnostic and Interventional Imaging · 2026
- Image quality and dose performance of photon-counting CT in multiple myeloma: a comparative clinical study.Journal of cancer research and clinical oncology · 2026
- Repeatability and reproducibility of spectral results with clinical photon-counting CT: multi-center phantom studymedRxiv · 2026
- Lumen imaging in calcified coronary arteries using high-resolution Gd-enhanced color K-edge imaging with spectral photon-counting CT: a phantom studyEuropean Radiology Experimental · 2026