Image quality robustness of HyperSight‐enabled versus conventional iterative cone‐beam CT across patient size and imaging geometry
Nicholas Nelson, Sorour Hosseini, Geoff Nelson, Thomas Martin, Bradley Nordin, Jessica Huang, Hui Zhao
Key points
From the paper's conclusion- Conclusions HS‐iCBCT demonstrated improved robustness to patient size and imaging geometry, providing superior image quality and HU stability compared to TB‐iCBCT.
Abstract
Background Recent advances in cone‐beam computed tomography (CBCT) have led to the development of next‐generation CBCT systems with enhanced image quality and quantitative imaging capabilities. These advances are particularly relevant in the context adaptive radiation therapy, where improved soft tissue visualization and Hounsfield unit (HU) stability are increasingly important for target localization and CBCT‐based dose calculation. Purpose To determine how image quality and HU stability change with increasing phantom size, non‐cylindrical geometries, and exposure level, and compare the robustness of Acuros‐based HyperSight‐enabled iterative CBCT (HS‐iCBCT) and conventional iterative CBCT on a TrueBeam (TB‐iCBCT) under these conditions. Methods The CatPhan® 604 phantom, combined with custom annular extensions, was used to simulate increasing patient size and non‐cylindrical imaging geometries. Three configurations (no annulus, small annulus, and large annulus) were imaged across a range of exposure levels (mAs) using HS‐iCBCT and TB‐iCBCT. Image evaluation metrics included noise, contrast‐to‐noise ratio (CNR), low‐contrast visibility (LCV), image uniformity, spatial resolution, and Hounsfield unit (HU) constancy. Results For the no‐annulus configuration, image quality was comparable between systems, with slightly lower noise observed for TB‐iCBCT. With increasing phantom size through the added annuli, TB‐iCBCT exhibited substantial degradation in noise, contrast, and uniformity, whereas HS‐iCBCT maintained more consistent performance. Spatial resolution was primarily dependent on field of view and was similar between systems, with smaller fields of view provided improved spatial resolution. HU constancy degraded with increasing phantom size and material density for both systems but remained substantially more stable for HS‐iCBCT. Conclusions HS‐iCBCT demonstrated improved robustness to patient size and imaging geometry, providing superior image quality and HU stability compared to TB‐iCBCT. These improvements may support potential lower‐dose imaging and more reliable performance in larger patients and non‐cylindrical imaging geometries commonly encountered in clinical radiotherapy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluating Image Quality of Cone Beam Computed Tomography Using HyperSight® on the Varian TrueBeam.Annals of biomedical engineering · 2026
- Towards adaptive radiotherapy: Comparing HyperSight and standard CBCT reconstructions on a C-Arm Linac.Journal of applied clinical medical physics · 2026
- Task-based image quality and dosimetric validation of HyperSight CBCT for adaptive radiotherapy.Journal of applied clinical medical physics · 2026
- HyperSight CBCT dose calculation accuracy for adaptive head and neck radiotherapy: Results from a prospective clinical trial.Medical physics · 2026
- A clinical evaluation of HyperSight CBCT enhanced offline adaptive breast radiotherapy.Frontiers in oncology · 2026
- Advanced HyperSight™ imaging for patients with adaptive SBRT of prostate cancer: a longitudinal analysis of tissue demarcationRadiation Oncology · 2025