Sex-based variability of water-equivalent diameter (Dw) and dosimetric parameters in adult head computed tomography (CT): implications for protocol optimisation.
Z Saga, A Rezzouk, C Anam, A Rahmouni
Laboratory of Solid-State Physics, Department of Physics, Faculty of Science Dhar El Mahraz, Sidi Mohamed Ben Abdellah University, Fez, Morocco. Electronic address: zouhir.saga@usmba.ac.ma. · Laboratory of Solid-State Physics, Department of Physics, Faculty of Science Dhar El Mahraz, Sidi Mohamed Ben Abdellah University, Fez, Morocco.
Key points
From the paper's conclusion- In this single-centre series, limited intersex variability in adult head CT dose metrics supports standardised protocols irrespective of sex. Dw is most valuable for dose audit, quality control, and identification of outlying examinations; its contribution to optimisation, which also requires image quality assessment, should be confirmed in prospective multicentre studies.
Abstract
AIM: The water-equivalent diameter (Dw) is increasingly used to express patient size in computed tomography (CT). Sex-related differences in Dw are documented in the chest and abdomen, but data for adult head CT remain scarce. The aims of this study were to compare Dw and dosimetric parameters (tube current-time product [mAs], volume computed tomography dose index [CTDIvol], size-specific dose estimates based on water-equivalent diameter [SSDEw], dose-length product [DLP], effective dose [EDw]) between men and women undergoing adult noncontrast head CT and to assess the relationship of Dw with mAs and SSDEw. MATERIALS AND METHODS: This retrospective study included 109 adult noncontrast head CT examinations (67 men and 42 women) performed with a standardised protocol on a 16-slice CT. Dw and dose indices were calculated using automated segmentation and compared between sexes (independent t-test); relationships were assessed by linear regression. RESULTS: No significant differences were found between sexes for Dw, mAs, CTDIvol, or SSDEw (all P > 0.05). DLP and EDw were slightly higher in men (P = 0.0337 and P = 0.0445), mainly owing to larger Dw and scan lengths. Dw correlated positively with mAs (women r = 0.57; men r = 0.46; P < 0.001) and SSDEw (women r = 0.52; men r = 0.38; P ≤ 0.002). CONCLUSION: In this single-centre series, limited intersex variability in adult head CT dose metrics supports standardised protocols irrespective of sex. Dw is most valuable for dose audit, quality control, and identification of outlying examinations; its contribution to optimisation, which also requires image quality assessment, should be confirmed in prospective multicentre studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Body Size, Radiation Dose and Exposure Index in Conventional Radiography: A Multicentre Prospective Study.Diagnostics (Basel, Switzerland) · 2026
- Adult diagnostic reference levels in computed tomography: A systematic review and analysis of regional variability.European journal of radiology · 2026
- Radiation Dose Assessment in Routine Adult Computed Tomography Examinations: A Multi-center Retrospective Study.Health physics · 2026
- Exploring the effectiveness of CARE Dose 4D and CARE kV Automatic Exposure Control Systems in CT: A Slim and Obese Phantom StudyPolish Journal of Medical Physics and Engineering · 2026
- Development of an Electronic Clinical Quality Measure for Excessive Radiation Dose or Inadequate Image Quality in Diagnostic CT in Adults.Journal of the American College of Radiology : JACR · 2026
- Doses in planning computed tomography in one radiotherapy department: Description of exposure parameters and estimation of organ doses, effective doses and typical doses.Radiography (London, England : 1995) · 2026