Computers in biology and medicine

A lightweight reliability- and confidence-aware selective ECG framework for degraded remote monitoring.

Nağme Bilgehan, Bülent Bilgehan

Imperial College Healthcare NHS Trust, London, United Kingdom. · Electrical and Electronic Engineering, Near East University, Mersin, Turkey; Near East University Science, Technology, Engineering Application and Research Center (BİLTEM), Mersin, Turkey. Electronic address: bulent.bilgehan@neu.edu.tr.

Key points

From the paper's conclusion
  • The results support a compact, degradation-aware selective ECG framework for edge- or gateway-assisted monitoring, while external wearable validation remains necessary.

Abstract

Remote electrocardiogram (ECG) monitoring in Internet of Medical Things (IoMT) systems is vulnerable to sensor noise, missing leads, packet loss, and temporal dropout. The proposed Lightweight Reliability-Aware Internet of Medical Things Electrocardiogram Network (LR-IoMT-ECGNet) is a lightweight reliability-aware framework that couples degradation-aware representation learning with selective decision support. Physikalisch-Technische Bundesanstalt eXtended Large (PTB-XL) electrocardiography dataset version 1.0.3 was used with official folds for training (n = 17,418), validation (n = 2183), and independent testing (n = 2198). Robustness was evaluated under clean, mild, moderate, and severe simulated degradation. A hybrid gate combining predictive confidence and signal reliability was tuned exclusively on validation data and locked before test evaluation. The 24,429-parameter model achieved an area under the receiver operating characteristic curve (AUC) of 0.9188 and F1-score of 0.8449 under clean conditions and an AUC of 0.8683, sensitivity of 0.7753, specificity of 0.8004, and F1-score of 0.8089 under severe degradation. By contrast, clean-trained CNN and ResNet-1D baselines showed near-zero severe specificity (0.0252 and 0.0143). On mixed-quality test data, the hybrid gate retained 90.2% of records and achieved accuracy of 0.8412, sensitivity of 0.8026, specificity of 0.8973, and F1-score of 0.8569 among accepted cases. Confidence-only referral was marginally stronger on some metrics, indicating that predictive confidence is the dominant selective signal while input reliability provides complementary information. The results support a compact, degradation-aware selective ECG framework for edge- or gateway-assisted monitoring, while external wearable validation remains necessary.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers