Interlayer-reinforced semi-IPN hydrogel coatings for durable lubrication and rivaroxaban-assisted antithrombotic activity on thermoplastic polyurethane.
Yuqi Zhang, Jiawang Yao, Dengdian Wang, Ke Cao, Sen Wang, Hongping Zhou
Key Laboratory of Structure and Functional Regulation of Hybrid Materials (Anhui University), Ministry of Education, School of Chemistry and Chemical Engineering, Anhui University, Hefei 230000, China. · Department of Urology, The Second Affiliated Hospital of Anhui Medical University, Hefei 230601, China.
Key points
From the paper's conclusion- Here, we developed an interlayer-reinforced semi-interpenetrating polymer network (semi-IPN) hydrogel coating for simultaneous lubrication and rivaroxaban-assisted antithrombotic modification of TPU.
Abstract
Thermoplastic polyurethane (TPU) is widely used in interventional medical devices because of its favorable mechanical properties and processability, but its inherent hydrophobicity and poor lubricity can promote tissue friction and thrombogenic responses during blood contact. Here, we developed an interlayer-reinforced semi-interpenetrating polymer network (semi-IPN) hydrogel coating for simultaneous lubrication and rivaroxaban-assisted antithrombotic modification of TPU. Polyvinylpyrrolidone (PVP) was physically interpenetrated within a crosslinked hydroxyethyl methacrylate/polyethylene glycol diacrylate (HEMA/PEGDA) network, while a polyurethane acrylate (PUA) interlayer was introduced to enhance coating-substrate interfacial stability. The optimized coating reduced the water contact angle of TPU from 91.10 ° to approximately 18.75 ° and decreased the coefficient of friction from 0.612 to 0.006, while maintaining stable lubrication over 50 friction cycles. Rivaroxaban was physically incorporated into the semi-IPN coating, and quantitative release measurements showed strong short-term drug retention, with only 0.413 ± 0.042% cumulative release in PBS after 24 h. The rivaroxaban-containing coating significantly prolonged activated partial thromboplastin time (APTT) to 67.33 ± 2.52 s, compared with 39.67 ± 1.53 s for the drug-free lubricious coating and 24.67 ± 1.15 s for unmodified TPU. Consistently, reduced platelet adhesion and thrombus formation were observed under static in vitro blood-contact conditions. Cytocompatibility evaluation further showed that the selected PUA-based coating maintained high cell viability, whereas PEI-based interlayers exhibited comparatively lower, although still acceptable, viability. These results demonstrate a robust interfacial strategy for constructing highly lubricious TPU surfaces with rivaroxaban-assisted antithrombotic functionality, while further long-term release, dynamic blood-contact, and in vivo evaluations remain necessary for translational assessment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Interface-Engineered PVA/PAA Composite Hydrogel Coatings on Bioinspired 3D-Printed PEEK Bilayers for Lubrication and Diclofenac DeliveryCoatings · 2026
- Polymeric Antithrombotic Coatings for Blood-Contacting Medical Devices: Evolution from Passive Protection to Adaptive RegulationPolymer Science & Technology · 2026
- Binary-Crosslinked Alginate/Zwitterionic Interpenetrating Polymer Network Coatings for Lubricious and Antifouling Urinary Catheters.ACS applied materials & interfaces · 2026
- Polysaccharide-Based Injectable Bioadhesive Hydrogel for Local Arterial Occlusion.ACS applied materials & interfaces · 2026
- Redox-active Polydopamine-copper network on hydrophilic silicone polymers for coupled antifouling and antimicrobial interfaces.Journal of colloid and interface science · 2026
- Advanced surface lubrication modifications for medical catheters: Materials, strategies, and evaluation.Advances in colloid and interface science · 2026