Clinical biomechanics (Bristol, Avon)

Biomechanical comparison of titanium cage versus three-dimensional-printed prosthesis for lumbar spinal tuberculosis reconstruction: a finite element analysis.

Pengpeng Ma, Xin Zhang, Zhiguo Zong, Su Liu, Wei Li, Chunling Zhang

Department of Orthopedics, The First Affiliated Hospital of Hebei North University, Zhangjiakou 075000, Hebei, China. · Department of Orthopedics, The First Affiliated Hospital of Hebei North University, Zhangjiakou 075000, Hebei, China. Electronic address: zhangchunlingzjk@sina.com.

Key points

From the paper's conclusion
  • Compared to titanium cage reconstruction, the 3D-printed custom prosthesis combined with posterior fixation provides biomechanically favorable performance, with lower stress concentrations in implants and bone, while maintaining equivalent segmental stability. These results suggest that 3DP may reduce risks of implant failure and subsidence in lumbar spinal tuberculosis reconstruction, although clinical validation is required.

Abstract

BACKGROUND: This finite element study aimed to compare the biomechanical performance of two reconstruction methods-pedicle screw fixation with a titanium mesh cage (TC) versus a patient-specific Three-Dimensional-printed porous prosthesis (3DP)-for treating L4-5 lumbar spinal tuberculosis. METHODS: A validated L1-S1 finite element model was developed from CT data of a healthy volunteer. Two surgical models were simulated: posterior fixation with a TC and with a customized 3DP. Under combined loading (400 N axial compression +7.5 N·m moments in flexion, extension, lateral bending, and rotation), range of motion (RoM), implant and vertebral bone stress (Von Mises), and adjacent segment intradiscal pressure were analyzed. FINDINGS: Both models effectively stabilized L4-5, with RoM below 1° in all directions. The 3DP model exhibited substantially lower peak implant stress (82.79 MPa vs. 477.9 MPa in flexion) and more uniform stress distribution. Vertebral bone stress adjacent to the 3DP exceeded the cortical bone yield threshold (90 MPa) only during flexion, whereas bone near the TC exceeded it in multiple directions. Stress in the pedicle screw system was also reduced in the 3DP model. No clinically meaningful difference in adjacent segment intradiscal pressure was observed between the two reconstructions. INTERPRETATION: Compared to titanium cage reconstruction, the 3D-printed custom prosthesis combined with posterior fixation provides biomechanically favorable performance, with lower stress concentrations in implants and bone, while maintaining equivalent segmental stability. These results suggest that 3DP may reduce risks of implant failure and subsidence in lumbar spinal tuberculosis reconstruction, although clinical validation is required.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers