Numerical modelling of a vascular palpation device for assessing intracranial aneurysm wall mechanics under heterogeneous conditions.
J Raviol, G Plet, R Hasegawa, K Yu, H Kosukegawa, M Ohta, H Magoariec, C Pailler-Mattei
Laboratoire de Tribologie et Dynamique des Systèmes, CNRS UMR 5513, Université de Lyon, École Centrale de Lyon, France. · Institute of Fluid Science, Tohoku University, 980-8577, Sendai Miyagi, Japan; Graduate School of Engineering, Tohoku University, 980-8579, Sendai Miyagi, Japan.
Key points
From the paper's conclusion- Within the modelling framework considered, inner wall displacements and luminal volume variation alone were inadequate to identify localized weakened zones, whereas the maximal principal strain appeared to be a pertinent quantity to detect and characterize such mechanical heterogeneities.
Abstract
BACKGROUND: The mechanical properties of the intracranial aneurysm wall are not taken into account in current rupture risk scores. A deformation device prototype was previously developed to characterize aneurysm wall mechanics in vivo through controlled deformation and inverse analysis. Its ability to detect localized mechanical heterogeneities within the aneurysm wall has not been assessed. METHODS: A finite element fluid-structure interaction model of a representative cerebral artery with a saccular aneurysm was developed. Seven mechanical models were considered: homogeneous walls, globally weakened aneurysm sacs, and models including localized weakened areas of 1 to 2.2 mm2 with rupture-associated parameters. Four device locations and two flow rates were studied under physiological blood flow conditions. Wall displacements, luminal volume variation, inner wall pressure and maximal principal strain were compared between configurations with and without the device. FINDINGS: Wall displacements induced by the device ranged from 372 to 733μm at systole peak, remaining within physiological bounds. Luminal volume variation did not allow the distinction between weakened areas and homogeneous walls. In contrast, the maximal principal strain variation showed an enlargement of ranges in weakened areas, and these ranges became distinguishable from homogeneous walls when the device was applied at systole against diastole without it. INTERPRETATION: Within the modelling framework considered, inner wall displacements and luminal volume variation alone were inadequate to identify localized weakened zones, whereas the maximal principal strain appeared to be a pertinent quantity to detect and characterize such mechanical heterogeneities.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Hemodynamic Differences Associated With Extreme Versus Minimal Proximal Neck Angulation in Ruptured Infrarenal Abdominal Aortic Aneurysms: A Patient-Specific Computational Study.Annals of vascular surgery · 2026
- Material characterization of heterogenous atherosclerotic arteries with the Virtual Fields Method.Computer methods and programs in biomedicine · 2026
- Hemodynamic-Structural Multi-Fidelity Assessment of Endovascular Flow Diversion Strategies for Cerebral Aneurysm Treatment: An Integrated Computational and Experimental FrameworkbioRxiv · 2026
- Local pressure as a dominant hemodynamic driver of wall enhancement in anterior communicating artery aneurysms: a facet-level computational fluid dynamics and vessel wall imaging analysis.Quantitative imaging in medicine and surgery · 2026
- Mechanical and Hemodynamic Effects of Balloon-Expandable and Self-Expanding Stents in a Saccular Abdominal Aortic Aneurysm Model: Fluid-Structure Interaction Analyses.Annals of vascular diseases · 2026
- A Fluid–Structure Interaction Computational Study of Residual Aortic Dissection to Investigate the Influence of Mechanical Behaviour of Wall and Flap on FlowsInternational Journal for Numerical Methods in Biomedical Engineering · 2025