Doping Control Center, Korea Institute of Science and Technology, Hwarang-ro 14-gil 5, Seongbuk-gu, Seoul, 02792, Republic of Korea; Department of Chemistry, Sogang University, Seoul, 04107, Republic of Korea. · Doping Control Center, Korea Institute of Science and Technology, Hwarang-ro 14-gil 5, Seongbuk-gu, Seoul, 02792, Republic of Korea; Department of Chemistry, Dongguk University, Seoul, 04620, Republic of Korea.
Key points
From the paper's conclusion- In the present study, background contamination of 39 environmental phenols originating from the HPLC system and sample-preparation materials and reagents was systematically evaluated, along with its impact on quantitative accuracy in human urine.
Abstract
Environmental phenols are ubiquitous environmental contaminants that frequently contaminate analytical systems and sample-preparation materials, thereby compromising quantitative accuracy. In the present study, background contamination of 39 environmental phenols originating from the HPLC system and sample-preparation materials and reagents was systematically evaluated, along with its impact on quantitative accuracy in human urine. Among the 39 environmental phenols investigated, bisphenol A, 4-nitrophenol (PNP), benzophenone-3 (BP-3), bisphenol AF (BPAF), and nonylphenol (NP) originated from the mobile phase and HPLC system. Contaminants originating from sample-preparation materials and reagents included PNP, BP-3, BPAF, methyl paraben, propyl paraben, 4-tert-octylphenol, and NP. Notably, the enzyme (β-glucuronidase/arylsulfatase from Helix pomatia) commonly used for urinary hydrolysis was associated with increased background responses for several analytes. Analysis of NIST SRMs 3672 and 3673 urine samples revealed that the influence of background contamination on quantitative accuracy depended on the specific analyte and its concentration. Furthermore, the correction strategies employed, namely blank subtraction and delay column, did not consistently perform across different analytes and concentrations. Consequently, neither strategy can be assumed to ensure optimal accuracy under all conditions. These findings underscore the need for a systematic assessment of background contamination and empirical validation of correction strategies to ensure the reliable quantification of environmental phenols in human urine, particularly when their concentrations are similar to background contamination levels.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Target quantification and ion mobility-HRMS suspect screening of BPA and its alternatives in human urine.Talanta · 2026
- Multi-class UHPLC-MS/MS method for the simultaneous determination of 32 endocrine-disrupting chemicals in the urine of pregnant women.Analytical and bioanalytical chemistry · 2026
- 48 Bicarbonate Measurement Discrepancies Between Blood Gas and Chemistry Analyzers: Role of Reagent Variability and LDH InterferenceAmerican Journal of Clinical Pathology · 2026
- Why Sensors Fail in Biological Samples: Fouling, Blocking, Matrix Effects and Prevention SolutionsInternational Journal of Molecular Sciences · 2026
- Assessing hair for human biomonitoring: comparison of organophosphorus flame-retardant levels in hair, urine and blood from the same individuals.Environment international · 2026
- Clearance and retention of leachables in ultrafiltration and diafiltration applications.International journal of pharmaceutics · 2026