Evaluation of the commutability of control materials for C-peptide proficiency testing.
M D Sooho Yu, Soye Jung, M D Yong Kwan Lim, M D Soo Jin Yoo, Hyung-Doo Park
Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea. · Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
Key points
From the paper's conclusion- Pooled human serum-based in-house prepared materials demonstrated more stable commutability than commercial materials, supporting their use in C-peptide EQA programs.
Abstract
BACKGROUND: Commutability refers to the consistency of measurement results between reference materials (RMs) and patient samples across different measurement procedures. C-peptide is essential for diabetes classification and management, however, inter-assay variability persists despite ongoing harmonization efforts. The use of commutable materials is therefore critical for calibration and external quality assessment (EQA). This study evaluated the commutability of pooled human serum and commercial control materials using CLSI- and IFCC-based approaches. METHODS: Commutability was evaluated using 30 clinical samples and six RMs, including three serum-based in-house prepared materials and three commercial materials. C-peptide was measured in triplicate using four major measurement procedures in Korea; Alinity i (Abbott Diagnostics), cobas e801 (Roche Diagnostics), UniCel DxI 800 (Beckman Coulter), and Atellica IM 1600 (Siemens Healthineers). Commutability was assessed using the CLSI EP14-ED4 Deming regression-based approach with 95% prediction intervals (PIs) and the IFCC difference-in-bias approach with expanded uncertainty, applying a total allowable error of 7.1% and a coverage factor of 1.9. RESULTS: Three clinical samples were excluded from the analyses after being identified as outliers in scatter plots. Using the CLSI approach with the remaining 27 clinical samples, all RMs fell within the 95% PIs across all method pairs and were classified as commutable. In contrast, the IFCC approach, with concentration partitioning at 1.0 ng/mL distinguished differences in commutability among the RMs. Serum-based in-house prepared materials were classified as commutable or indeterminate, whereas commercial materials showed multiple indeterminate results and one non-commutable result. CONCLUSION: Pooled human serum-based in-house prepared materials demonstrated more stable commutability than commercial materials, supporting their use in C-peptide EQA programs.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Commutability Assessment of Materials for Serum Copper Measurement Using the IFCC Difference-in-Bias and the CLSI EP30 Approaches.Annals of laboratory medicine · 2026
- Commonly used C‐peptide assays show differing associations with CGM metricsDiabetes Obesity and Metabolism · 2025
- Evaluating feasibility of ferritin assay harmonisation with commutable proof-of-concept reference materials and EQA-based monitoring.Clinical chemistry and laboratory medicine · 2026
- Strengthening the interpretation of longitudinal EQA data through the sporadic inclusion of native human sera in general clinical chemistry.Frontiers in molecular biosciences · 2026
- The second assessment of the commutability of candidate reference materials for the harmonization of blood neurofilament light chain measurements.Clinical chemistry and laboratory medicine · 2026
- Commutability assessment of candidate reference materials for total bilirubin assays in neonates.Clinical chemistry and laboratory medicine · 2026