Cardiovascular revascularization medicine : including molecular interventions

Stepwise intracoronary imaging assessment of lumen and vessel changes during drug-coated balloon angioplasty for in-stent restenosis.

Georgios M Alexandridis, Jari A van der Eijk, Karol Sadowski, Isabella Kardys, Wijnand K den Dekker, Roberto Diletti, Rutger-Jan Nuis, Karim D Mahmoud,

Department of (Interventional) Cardiology, Thoraxcenter, Erasmus University Medical Center, Rotterdam, the Netherlands. · Department of (Interventional) Cardiology, Thoraxcenter, Erasmus University Medical Center, Rotterdam, the Netherlands; Biomedical Instrument Institute, School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, China.

Observational · n=20

Key points

From the paper's conclusion
  • ICI demonstrates that the principal morphologic changes are achieved after lesion preparation, with smaller yet significant additional acute mechanical effects following DCB treatment.

Abstract

BACKGROUND: Drug-coated balloon (DCB) angioplasty is an established treatment for in-stent restenosis (ISR), and its procedural success mainly depends on optimal lesion preparation. Angiography alone offers limited assessment of acute procedural effects, and sequential vessel and lumen changes have not been comprehensively characterized with intracoronary imaging (ICI). Therefore, we aimed to characterize the stepwise vessel and lumen changes during DCB angioplasty for ISR across three procedural stages: pre-lesion preparation, post-lesion preparation and post-DCB inflation. METHODS: We retrospectively enrolled 20 patients (21 lesions) undergoing DCB angioplasty for ISR with available ICI pullbacks (intravascular ultrasound (IVUS), n = 16; optical coherence tomography (OCT), n = 5) at three stages: pre-lesion preparation, post-lesion preparation, and post-DCB inflation. Lesions treated with DCBs were identified on angiography and co-registered to the ICI pullbacks. RESULTS: Median age was 69 years, 15 patients (75%) were male, and lesion preparation was performed with non-compliant balloons in 76%. On IVUS, minimal lumen area (MLA) increased from 2.7 [2.2-3.5]mm2 at baseline to 4.8 [4.1-6.0]mm2 after lesion preparation (median difference (95% confidence interval) +2.6 (1.8-2.7)mm2, p < 0.001), and 5.9 [4.5-7.0]mm2 post-DCB (+0.6 (0.4-1.1)mm2, p < 0.001). On OCT, MLA increased from 1.9 [1.4-2.1]mm2 at baseline to 3.8 [2.9-4.2]mm2 after lesion preparation (+2.0 [0.3-2.8]mm2), and 4.5 [3.9-5.3]mm2 post-DCB (+0.6 [0.3-1.0]mm2). Neointimal burden decreased after lesion preparation and decreased further after DCB treatment. CONCLUSION: ICI demonstrates that the principal morphologic changes are achieved after lesion preparation, with smaller yet significant additional acute mechanical effects following DCB treatment.

Read the paper