Short-frame, self-expanding, fully recapturable transcatheter aortic valve replacement device in pure aortic regurgitation: Procedural performance and 30-day outcomes.
Hui Yan, Yunyao Li, Weiguo Wan, Haihan Liao, Sheng Cao, Hongning Song, Xiaoya Zhou, Hong Jiang,
Department of Cardiology, Renmin Hospital of Wuhan University, Wuhan, Hubei, 430060, China; Hubei Key Laboratory of Cardiology, Wuhan, Hubei, 430000, China; Cardiovascular Research Institute, Wuhan University, Wuhan, Hubei, 430000, China. · Department of Cardiology, Renmin Hospital of Wuhan University, Wuhan, Hubei, 430060, China.
Key points
From the paper's conclusion- In this exploratory single-centre series, transfemoral TAVR using a short-frame, fully recapturable self-expanding valve achieved 100% technical success and favourable early safety and haemodynamic performance in selected patients with PNAR. Prospective multicentre studies with longer follow-up and appropriate comparator groups are warranted.
Abstract
BACKGROUND: Achieving stable anchoring and effective sealing remains the principal limitation of transcatheter aortic valve replacement (TAVR) in patients with pure native aortic regurgitation (PNAR). A short-frame, fully recapturable self-expanding valve may expand the procedural margin by enabling iterative repositioning and implantation depth optimisation in non-calcified anatomy. METHODS: We analysed 13 consecutive patients with symptomatic moderate-to-severe or severe PNAR who underwent transfemoral TAVR with the ScienCrown® (Lepu Medical, Beijing, China) system. The primary endpoint was technical success at exit from the procedure room according to VARC-3 criteria. Secondary endpoints included device success and early safety at 30 days. RESULTS: Technical success was achieved in all patients (13/13, 100%), and device success at 30 days was also achieved in all patients (13/13, 100%). The 30-day early safety endpoint was met in 12 patients (92.3%). The mean prosthesis size was 27.62 ± 1.26 mm, with a mean prosthesis diameter oversizing relative to the average annular diameter of 12.90 ± 3.54%; no balloon predilatation was performed. The mean number of deployment attempts was 1.62 ± 0.77. All procedures were completed without valve embolisation/migration, coronary obstruction, second-valve implantation, or emergency surgery. One patient (7.69%) developed complete atrioventricular block on postoperative day 3 and underwent permanent pacemaker implantation; no additional pacemaker implantations occurred through 30 days. Predischarge diffusion-weighted magnetic resonance imaging demonstrated no new acute cerebral infarction. At 30 days, there were no deaths, strokes or transient ischaemic attacks, myocardial infarction, life-threatening/major bleeding, major vascular complications, endocarditis, or valve-related reinterventions. Mean transvalvular gradients were 7.67 ± 5.20 mmHg, and all patients had improved or stable NYHA functional class with no moderate or severe residual aortic regurgitation. CONCLUSIONS: In this exploratory single-centre series, transfemoral TAVR using a short-frame, fully recapturable self-expanding valve achieved 100% technical success and favourable early safety and haemodynamic performance in selected patients with PNAR. Prospective multicentre studies with longer follow-up and appropriate comparator groups are warranted.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Transcatheter Aortic Valve Implantation for Pure Native Aortic Regurgitation: A Comprehensive State-of-the-Art Review.Reviews in cardiovascular medicine · 2026
- Dedicated versus non-dedicated transcatheter valves for pure native aortic regurgitation: a single-centre experienceNetherlands Heart Journal · 2026
- Ascending aorta and pure aortic regurgitation transcatheter aortic valve replacement: insights from the National Transcatheter Valve Therapeutics Registry.Heart (British Cardiac Society) · 2026
- Clinical Outcomes of a Dedicated TAVR Device for Native Aortic Regurgitation: Insights From a Multicenter Real-World Experience.Circulation. Cardiovascular interventions · 2026
- Transcatheter Aortic Valve Replacement in Patients With Symptomatic, Severe Aortic Regurgitation: 1-Year Outcomes of the J-Valve Transfemoral Early Feasibility Study.Journal of the Society for Cardiovascular Angiography & Interventions · 2026
- Dedicated versus off-label transcatheter devices for pure native aortic regurgitation via transfemoral access: a meta-analysis of contemporary outcomes.AsiaIntervention · 2026