Monte Carlo characterization study of a TLD-600/TLD-700 neutron personal dosimeter exposed to a 252Cf source on an ISO water phantom at LPN-CIEMAT.
Lenin E Cevallos-Robalino, Roberto Méndez-Villafañe, Hector Rene Vega-Carrillo, Marco Aurélio de Sousa Lacerda, Víctor David Larco-Torres, Orlando Barcia-Ayala
Universidad Politécnica Salesiana (UPS), Carrera de Ingeniería en Mecatrónica, Grupo de Investigación en Micro-Nanotecnología y Energía Nuclear (NANOTECH), C. Robles 107 Chambers, 090108, Guayas, Guayaquil, Ecuador; Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas (CIEMAT), Laboratorio de Patrones Neutrónicos, Avenida complutense 40, 28040, Madrid, Spain. Electronic address: lcevallos@ups.edu.ec. · Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas (CIEMAT), Laboratorio de Patrones Neutrónicos, Avenida complutense 40, 28040, Madrid, Spain.
Key points
From the paper's conclusion- This study presents a detailed Monte Carlo characterization of a neutron personal dosimeter based on thermoluminescent detectors (TLD-600 and TLD-700), irradiated with a252Cf neutron source on an ISO water phantom at the Neutron Standards Laboratory (LPN-CIEMAT, Spain).
Abstract
This study presents a detailed Monte Carlo characterization of a neutron personal dosimeter based on thermoluminescent detectors (TLD-600 and TLD-700), irradiated with a252Cf neutron source on an ISO water phantom at the Neutron Standards Laboratory (LPN-CIEMAT, Spain). A set of six dosimeters, modeled using the MCNP6 code, was positioned on the phantom surface and exposed to the neutron field at various source-to-detector distances: 50, 75, 100, and 125 cm. The simulations included both simplified (vacuum) and realistic configurations that considered all laboratory structures-walls, bench, and pool-to evaluate the influence of backscattered neutrons on detector response. The TLD-600/TLD-700 pairs, with and without cadmium (Cd) filters, were analyzed to distinguish thermal and fast neutron contributions. The results showed a clear decrease in neutron response with increasing distance from the source and confirmed the Cd filter's effectiveness in suppressing thermal neutrons. Additional figures introduced in this version provide detailed visualizations of the laboratory geometry, dosimeter configuration, neutron flux distribution, and comparative response profiles, enhancing the understanding of spatial effects and scattering contributions. The main contribution of this work is the facility-specific computational assessment of the influence of laboratory structures on the neutron response of TLD-600/TLD-700 personal dosimeters under a252Cf reference field. By explicitly modeling the irradiation room, water pool, and supporting structures, the simulations quantify the contribution of environmental neutron backscattering, which produced an increase of approximately 5-8% in the calculated TLD-600 reaction rates compared with the vacuum configuration. The analysis also evaluates the effects of source-to-detector distance and cadmium filtering on the simulated detector response. These results provide facility-specific information relevant to neutron field characterization and support future experimental calibration studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Experimental validation of ambient dose equivalent rate calculations for a 241Am-Be neutron source using OpenMC code.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Characterization of the fast neutron response of liquid scintillator detectors at NFSarXiv · 2026
- Monte Carlo simulations of a custom-made neutron probe based on the SI-19N proportional 3He counter.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- A comprehensive microdosimetric study of neutron capture enhanced proton therapy: physical dose to RBE-based evaluation.Journal of radiation research · 2026
- Experimental validation of a newly developed neutron ambient dose H* detector in an accelerator-based BNCT fieldJapanese Journal of Applied Physics · 2026
- Nuclear and radiological characterization of a 226Ra-Be neutron irradiator at INFISA-UNMSM, Peru, using Monte Carlo simulation with the OpenMC codeNuclear Engineering and Technology · 2026